Significant Radiation Dose Reduction in the Hybrid Operating Room Using a Novel X-ray Imaging Technology.

R F F van den Haak1, B C Hamans2, K Zuurmond3, B A N Verhoeven4, O H J Koning4.   

Abstract

OBJECTIVE/
BACKGROUND: To prospectively quantify radiation dose change in aortoiliac endovascular procedures in the hybrid operating room (OR) for patients and medical staff with a novel X-ray imaging technology (ClarityIQ technology), and to assess whether procedure or fluoroscopy time or dose of iodinated contrast was affected.
METHODS: A prospective study including 138 patients was performed to compare radiation dose before and after installation of a novel X-ray imaging technology. Endovascular aneurysm repair (EVAR) was performed in 37 patients and an endovascular procedure for aortoiliac occlusive disease (AIOD) in 101. Patient radiation dose in air kerma (AK) and dose area product (DAP), patient demographics, and procedural data were recorded. Staff radiation dose was measured with real time personal dosimetry measurements. In both the EVAR and AIOD groups the reference system, ALX (AlluraXper FD20; Philips Healthcare, Best, the Netherlands), was compared with the upgraded X-ray system, CIQ (AlluraClarity FD20; Philips Healthcare). Procedure time, fluoroscopy time, and iodinated contrast dose were recorded.
RESULTS: Patient radiation dose reduction in the EVAR group, in median AK, was 56% (ALX = 1,262.5 mGy; CIQ = 556.0 mGy [p < .01]); and in median DAP it was 57% (ALX = 224.4 Gycm(2) and CIQ = 95.8 Gycm(2) [p < .01]). Patient radiation dose reduction in the AIOD group, in median AK, was 76% (ALX = 1,011.0 mGy; CIQ = 248.0 mGy [p < .01]); and in median DAP it was 73% (ALX = 138.1 Gycm(2); CIQ = 38.0 Gycm(2) [p < .01]). Staff dose reduction in the EVAR group was 16% (ALX = 70.1 μSv; CIQ = 59.2 μSv [p = .43]) and in the AIOD group it was 69% (ALX = 96.2 μSv; CIQ = 30.1 μSv [p < .01]). There was no statistically significant difference between patient demographics, procedure time, fluoroscopy time, and iodinated contrast medium use in the two treatment groups before and after installation.
CONCLUSION: A novel X-ray imaging technology in the hybrid OR suite resulted in a significant reduction of patient and staff radiation dose without affecting procedure length, fluoroscopy time, or use of contrast.
Copyright © 2015 European Society for Vascular Surgery. Published by Elsevier Ltd. All rights reserved.

Entities:  

Keywords:  Aneurysm repair; Dosimetry; Endovascular; Radiation; Radiation exposure

Mesh:

Substances:

Year:  2015        PMID: 26286386     DOI: 10.1016/j.ejvs.2015.06.025

Source DB:  PubMed          Journal:  Eur J Vasc Endovasc Surg        ISSN: 1078-5884            Impact factor:   7.069


  8 in total

Review 1.  [Hybrid operating theater].

Authors:  T Fuchs-Buder; N Settembre; D Schmartz
Journal:  Anaesthesist       Date:  2018-07       Impact factor: 1.041

Review 2.  Physician and Patient Radiation Exposure During Endovascular Procedures.

Authors:  Andrew M Goldsweig; J Dawn Abbott; Herbert D Aronow
Journal:  Curr Treat Options Cardiovasc Med       Date:  2017-02

3.  Reduced Patient Radiation Exposure during Neurodiagnostic and Interventional X-Ray Angiography with a New Imaging Platform.

Authors:  K van der Marel; S Vedantham; I M J van der Bom; M Howk; T Narain; K Ty; A Karellas; M J Gounis; A S Puri; A K Wakhloo
Journal:  AJNR Am J Neuroradiol       Date:  2017-01-19       Impact factor: 3.825

4.  Radiation dose during endovascular aneurysm repair (EVAR): upgrade of an angiographic system from standard to Eco mode.

Authors:  Maria Antonella Ruffino; Marco Fronda; Andrea Discalzi; Paola Isoardi; Laura Bergamasco; Roberto Ropolo; Dorico Righi; Paolo Fonio
Journal:  Radiol Med       Date:  2018-07-23       Impact factor: 3.469

5.  Impact of dose reducing software on patient and staff temple dose during fluoroscopically guided pacemaker insertion, closure devices implantation and coronary angiography procedures.

Authors:  Kelly S Wilson-Stewart; Davide Fontanarosa; Eva Malacova; Jamie V Trapp
Journal:  Phys Eng Sci Med       Date:  2022-05-09

6.  Radiation Awareness for Endovascular Abdominal Aortic Aneurysm Repair in the Hybrid Operating Room. An Instant Patient Risk Chart for Daily Practice.

Authors:  Quirina M de Ruiter; Crystel M Gijsberts; Constantijn E Hazenberg; Frans L Moll; Joost A van Herwaarden
Journal:  J Endovasc Ther       Date:  2017-04-10       Impact factor: 3.487

7.  Radiation dose reference card for interventional radiology procedures: Experience in a tertiary referral centre.

Authors:  Anna Varghese; Shyamkumar N Keshava; Vinu Moses; George Koshy; Suraj Mammen; Munawwar Ahmed; Roshan S Livingstone
Journal:  Indian J Radiol Imaging       Date:  2019-10-30

8.  Radiation exposure for intraoperative 3D scans in a hybrid operating room: how to reduce radiation exposure for the surgical team.

Authors:  K Schuetze; M Kraus; A Eickhoff; F Gebhard; P H Richter
Journal:  Int J Comput Assist Radiol Surg       Date:  2018-03-29       Impact factor: 2.924

  8 in total