In Vitro Comparison of Active and Passive Physiological Control Systems for Biventricular Assist Devices.

Jo P Pauls1,2, Michael C Stevens3,4, Emma Schummy3, Geoff Tansley5,3, John F Fraser3,6, Daniel Timms7,8, Shaun D Gregory3,6.   

Abstract

The low preload and high afterload sensitivities of rotary ventricular assist devices (VADs) may cause ventricular suction events or venous congestion. This is particularly problematic with rotary biventricular support (BiVAD), where the Starling response is diminished in both ventricles. Therefore, VADs may benefit from physiological control systems to prevent adverse events. This study compares active, passive and combined physiological controllers for rotary BiVAD support with constant speed mode. Systemic (SVR) and pulmonary (PVR) vascular resistance changes and exercise were simulated in a mock circulation loop to evaluate the capacity of each controller to prevent suction and congestion and increase exercise capacity. All controllers prevented suction and congestion at high levels of PVR (900 dynes s cm(-5)) and SVR (3000 dynes s cm(-5)), however these events occurred in constant speed mode. The controllers increased preload sensitivity (0.198-0.34 L min(-1) mmHg(-1)) and reduced afterload sensitivity (0.0001-0.008 L min(-1) mmHg(-1)) of the VADs when compared to constant speed mode (0.091 and 0.072 L min(-1) mmHg(-1) respectively). The active controller increased pump speeds (400-800 rpm) and pump flow by 2.8 L min(-1) during exercise, thus increasing exercise capacity. By reducing suction and congestion and by increasing exercise capacity, the control systems presented in this study may help increase quality of life of VAD patients.

Entities:  

Keywords:  Active control; Passive control; Physiological control; Ventricular assist devices; Ventricular suction prevention

Mesh:

Year:  2015        PMID: 26283049     DOI: 10.1007/s10439-015-1425-1

Source DB:  PubMed          Journal:  Ann Biomed Eng        ISSN: 0090-6964            Impact factor:   3.934


  1 in total

1.  Preload Sensitivity with TORVAD Counterpulse Support Prevents Suction and Overpumping.

Authors:  Jeffrey R Gohean; Erik R Larson; Raul G Longoria; Mark Kurusz; Richard W Smalling
Journal:  Cardiovasc Eng Technol       Date:  2019-06-11       Impact factor: 2.495

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.