Measurement of the frequency response and common-mode gain of neonatal respiratory pressure and flow measurement systems. Part 2: Results.

M J Turner1, I M Macleod, A D Rothberg.   

Abstract

We show that the technique of exciting a differential pressure transducer at one port to measure its differential gain yields incorrect results in the case of asymmetrical differential pressure transducers but is acceptable in the case of symmetrical transducers. We have measured the common-mode gain of a symmetrical and an asymmetrical differential pressure transducer from 1 Hz to 100 Hz, both directly and by computation from differential gain measurements. No compensation for common-mode error is necessary when a symmetrical transducer is used in mechanically ventilated neonates. We have also measured the frequency response of neonatal Fleisch and screen pneumotachographs connected to Validyne MP45 differential pressure transducers in air, 60% oxygen and 100% oxygen, and concluded that the effect of oxygen concentration is small below 40 Hz. However, the normalised frequency response of the flow transducer differs markedly from unity at frequencies inside the flow bandwidth generated by neonatal pressure cycled ventilators and dynamic correction is therefore necessary.

Entities:  

Mesh:

Year:  1989        PMID: 2627765     DOI: 10.1088/0143-0815/10/3/003

Source DB:  PubMed          Journal:  Clin Phys Physiol Meas        ISSN: 0143-0815


  2 in total

1.  Accuracy of deadspace free ventilatory measurements for lung function testing in ventilated newborns: a simulation study.

Authors:  B Foitzik; P Schaller; M Schmidt; G Schmalisch
Journal:  J Clin Monit Comput       Date:  2000       Impact factor: 2.502

2.  Two-point calibration procedure of the forced oscillation technique.

Authors:  K N Desager; M Cauberghs; K P Van de Woestijne
Journal:  Med Biol Eng Comput       Date:  1997-11       Impact factor: 2.602

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.