| Literature DB >> 26269721 |
Peta Ellen Tehan1, Vivienne Helaine Chuter1.
Abstract
BACKGROUND: Hand held Doppler examination is a frequently used non-invasive vascular assessment utilised by podiatrists. Despite this, the reliability of hand-held Doppler has not been thoroughly investigated. Given the importance of Doppler in completing a vascular assessment of the lower limb, it is essential to determine the reliability of the interpretation of this testing method in practicing podiatrists.Entities:
Keywords: Doppler; Peripheral arterial disease; Reliability
Year: 2015 PMID: 26269721 PMCID: PMC4534124 DOI: 10.1186/s13047-015-0097-2
Source DB: PubMed Journal: J Foot Ankle Res ISSN: 1757-1146 Impact factor: 2.303
Fig. 1Flow diagrams for the different components of the study
Participant characteristics
| Public participants | Private participants | Audio interpretation | |
|---|---|---|---|
| Males n (%) | 17 (53) | 18 (58) | 17 (56) |
| Females n (%) | 15 (47) | 13 (42) | 13 (44) |
| Mean age (years) | 70.9 (SD 7.1) | 72.0 (SD 5.7) | 71.6 (SD 6.7) |
| Age range (years) | 57 - 88 | 61 - 81 | 55 - 82 |
| DM n (%) | 23 (72) | 15 (48) | 19 (63) |
| Total N | 32 | 31 | 30 |
For Doppler use the public participant group was evaluated by the public practice raters, and private participants were evaluated by private practice raters. For visual Doppler waveform analysis, a sub-set of 30 printed waveforms from both public and private participants were randomly selected and evaluated by all raters. For audio interpretation all raters evaluated the recorded sounds of the sub-group listed above
SD standard deviation, DM diabetes mellitus
Reliability results for use of Doppler
| Inter-rater reliability | Intra-rater reliability | ||||||||
|---|---|---|---|---|---|---|---|---|---|
| DP | 95 % CI | PT | 95 % CI | DP | 95 % CI | PT | 95 % CI | ||
| Private | κ 0.20 ( | −0.09 to 0.49 | κ 0.16 ( | −0.11 to 0.43 | Private | κ 0.22 ( | −0.31 to 0.53 | κ 0.42 ( | 0.15 to 0.69 |
| Public | κ 0.17 ( | −0.14 to 0.48 | κ 0.24 ( | −0.07 to 0.55 | Public | κ 0.21 ( | −0.16 to 0.58 | κ 0.27 ( | −0.06 to 0.60 |
95 % CI 95 % confidence intervals, DP dorsalis pedis artery, PT posterior tibial artery, Private private practitioners, Public public practitioners
Reliability results for Doppler audio interpretation
| Inter-rater reliability | 95 % CI | Intra-rater reliability | 95 % CI | ||
|---|---|---|---|---|---|
| Private | κ 0.31 ( | −0.08 to 0.70 | Private | κ 0.53 ( | 0.16 to 0.91 |
| Public | κ 0.61 ( | 0.23 to 0.99 | Public | κ 1.00 ( | 1.00 to 1.00 |
95 % CI 95 % confidence intervals, Private private practitioners, Public public practitioners
Reliability results for Doppler visual interpretation
| Inter-rater reliability | 95 % CI | Intra-rater reliability | 95 % CI | ||
|---|---|---|---|---|---|
| Private | κ 0.90 ( | 0.71 to 1.09 | Private | κ 1.00 ( | 1.00 to 1.00 |
| Public | κ 0.77 ( | 0.53 to 1.01 | Public | κ 0.81 ( | 0.57 to 1.05 |
95 % CI 95 % confidence intervals, Private private practitioners, Public public Practitioners