Literature DB >> 26265071

Apical closure device for transapical valve procedures.

Enrico Ferrari1, Denis Berdajs2, Piergiorgio Tozzi3, Giuseppe Siniscalchi3, Ludwig K von Segesser2.   

Abstract

OBJECTIVES: Transapical transcatheter valve procedures are performed through a left minithoracotomy and require apical sutures to seal the apical access site. The use of large-calibre devices compromises any attempt to fully perform the procedure with a thoracoscopic approach or percutaneously. We report our preliminary experience in animals with a new sutureless self-expandable apical occluder, engineered to perform transapical access site closure in a minimally invasive setting with large-size introducer sheaths.
METHODS: The apical occluder with extendable waist was implanted in six young pigs during an acute animal study. Under general anaesthesia, animals (mean weight: 62 ± 8 kg) received full heparinization (heparin: 100 UI/kg; activated clotting time above 250 s). Through a median sternotomy, a 21-Fr Certitude™ introducer sheath (outer diameter: 25 Fr) was placed over the wire into the cardiac apex. The delivery catheter carrying the constrained apical plug was inserted into the sheath and deployed under fluoroscopic control, whereas the Certitude™ was retrieved. After protamine infusion, we observed and recorded the 1-h bleeding with standard haemodynamic parameters. Animals were sacrificed, and hearts analysed.
RESULTS: Six apical closure devices were successfully introduced and deployed in six pig hearts through large-size apical sheaths at first attempt. In all animals, the plugs guaranteed immediate apical sealing and traces of blood were collected in the pericardium during the 1-h observational period (mean of 16 ± 3.4 ml of blood loss per animal). Haemodynamic parameters remained stable during the entire study period and no plug dislodgement was detected with normal systemic blood pressure (mean arterial mean blood pressure: 65 ± 7 mmHg). Post-mortem analysis confirmed the full deployment and good fixation of all plugs, without macroscopic damages to the surrounding myocardium.
CONCLUSIONS: This sutureless self-expandable apical occluder is a simple device capable of sealing large-size apical access sites (20-35 Fr) in an acute animal study. This approach is a step further towards less invasive transapical valve procedures in the clinical setting, and further animal tests will be performed to confirm the long-term efficacy and safety of this device.
© The Author 2015. Published by Oxford University Press on behalf of the European Association for Cardio-Thoracic Surgery. All rights reserved.

Entities:  

Keywords:  Closure device; Percutaneous heart procedures; Transapical valve replacement; Transcatheter aortic valve replacement

Mesh:

Year:  2015        PMID: 26265071     DOI: 10.1093/icvts/ivv225

Source DB:  PubMed          Journal:  Interact Cardiovasc Thorac Surg        ISSN: 1569-9285


  1 in total

1.  Utility of dual longitudinal diameter-reducing ties in aortic arch thoracic endovascular aortic repair.

Authors:  Arkadiusz Kazimierczak; Paweł Rynio; Tomasz Jędrzejczak; Maciej Lewandowski; Piotr Gutowski
Journal:  Postepy Kardiol Interwencyjnej       Date:  2019-12-08       Impact factor: 1.426

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.