Balancing Radiation and Contrast Media Dose in Single-Pass Abdominal Multidetector CT: Prospective Evaluation of Image Quality.

Luigi Camera1, Federica Romano2, Immacolata Liccardo2, Raffaele Liuzzi3, Massimo Imbriaco2, Pier Paolo Mainenti3, Laura Micol Pizzuti3, Sabrina Segreto2, Simone Maurea2, Arturo Brunetti2.   

Abstract

RATIONALE AND
OBJECTIVES: As both contrast and radiation dose affect the quality of CT images, a constant image quality in abdominal contrast-enhanced multidetector computed tomography (CE-MDCT) could be obtained balancing radiation and contrast media dose according to the age of the patients.
MATERIALS AND METHODS: Seventy-two (38 Men; 34 women; aged 20-83 years) patients underwent a single-pass abdominal CE-MDCT. Patients were divided into three different age groups: A (20-44 years); B (45-65 years); and C (>65 years). For each group, a different noise index (NI) and contrast media dose (370 mgI/mL) was selected as follows: A (NI, 15; 2.5 mL/kg), B (NI, 12.5; 2 mL/kg), and C (NI, 10; 1.5 mL/kg). Radiation exposure was reported as dose-length product (DLP) in mGy × cm. For quantitative analysis, signal-to-noise (SNR) and contrast-to-noise (CNR) ratios were calculated for both the liver (L) and the abdominal aorta (A). Statistical analysis was performed with a one-way analysis of variance. Standard imaging criteria were used for qualitative analysis.
RESULTS: Although peak hepatic enhancement was 152 ± 16, 128 ± 12, and 101 ± 14 Hounsfield units (P < .001) for groups A, B, and C, respectively, no significant differences were observed in the corresponding SNRL with 9.2 ± 1.4, 9.1 ± 1.2, and 9.2 ± 3. Radiation (mGy × cm) and contrast media dose (mL) administered were 476 ± 147 and 155 ± 27 for group A, 926 ± 291 and 130 ± 16 for group B, and 1981 ± 451 and 106 ± 15 for group C, respectively (P < .001). None of the studies was graded as poor or inadequate by both readers, and the prevalence-adjusted bias-adjusted kappa ranged between 0.48 and 0.93 for all but one criteria.
CONCLUSIONS: A constant image quality in CE-MDCT can be obtained balancing radiation and contrast media dose administered to patients of different age.
Copyright © 2015 AUR. Published by Elsevier Inc. All rights reserved.

Entities:  

Keywords:  CT protocols; CT technique; contrast media; effective dose; image quality

Mesh:

Substances:

Year:  2015        PMID: 26264765     DOI: 10.1016/j.acra.2015.06.016

Source DB:  PubMed          Journal:  Acad Radiol        ISSN: 1076-6332            Impact factor:   3.173


  1 in total

1.  Diagnostic efficacy of single-pass abdominal multidetector-row CT: prospective evaluation of a low dose protocol.

Authors:  Luigi Camera; Immacolata Liccardo; Federica Romano; Raffaele Liuzzi; Antonio Rispo; Massimo Imbriaco; Anna Testa; Gaetano Luglio; Simona De Fronzo; Fabiana Castiglione; Luigi Bucci; Arturo Brunetti
Journal:  Br J Radiol       Date:  2016-11-09       Impact factor: 3.039

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.