| Literature DB >> 26263060 |
Luyan Zhang1, Wei Zhang1, Gang Chen2.
Abstract
This report describes the fabrication and the application of a novel carbon nanotube (CNT)-epoxy composite electrode as a sensitive amperometric detector for the capillary electrophoresis (CE). The composite electrode was fabricated on the basis of the in situ polycondensation of a mixture of CNTs and 1,2-ethanediamine-containing bisphenol A epoxy resin in the inner bore of a piece of fused silica capillary under heat. It was coupled with CE for the separation and detection of arbutin and bergenin in Bergeniae Rhizoma, a traditional Chinese medicine, to demonstrate its feasibility and performance. The two phenolic constituents were well separated within 10min in a 45cm capillary length at a separation voltage of 12kV using a 50mM borate buffer (pH 9.2). The CNT-based detector offered higher sensitivity, significantly lower operating potential, satisfactory resistance to surface fouling, and lower expense of operation, indicating great promise for a wide range of analytical applications. It showed long-term stability and reproducibility with relative standard deviations of less than 5% for the peak current (n=15).Entities:
Keywords: Amperometric detection; Arbutin; Bergenin; Capillary electrophoresis; Carbon nanotube; Epoxy
Mesh:
Substances:
Year: 2015 PMID: 26263060 DOI: 10.1016/j.jpba.2015.07.044
Source DB: PubMed Journal: J Pharm Biomed Anal ISSN: 0731-7085 Impact factor: 3.935