Yutao Jiang1, Daniel Hascall2, Delia Li2, Joseph H Pease2.
Abstract
In this paper, we introduce a high throughput LCMS/UV/CAD/CLND system that improves upon previously reported systems by increasing both the quantitation accuracy and the range of compounds amenable to testing, in particular, low molecular weight "fragment" compounds. This system consists of a charged aerosol detector (CAD) and chemiluminescent nitrogen detector (CLND) added to a LCMS/UV system. Our results show that the addition of CAD and CLND to LCMS/UV is more reliable for concentration determination for a wider range of compounds than either detector alone. Our setup also allows for the parallel analysis of each sample by all four detectors and so does not significantly increase run time per sample.Entities:
Keywords: Charged aerosol detector; Chemiluminescent nitrogen detector; Fragment; Liquid chromatography; Pharmaceutical; Purity
Mesh:
Substances:
Year: 2015 PMID: 26256922 DOI: 10.1016/j.chroma.2015.07.112
Source DB: PubMed Journal: J Chromatogr A ISSN: 0021-9673 Impact factor: 4.759