| Literature DB >> 26251906 |
Francisco J Casas1, David Ortiz2, Enrique Villa3, Juan L Cano4, Jaime Cagigas4, Ana R Pérez5, Beatriz Aja6, J Vicente Terán7, Luisa de la Fuente8, Eduardo Artal9, Roger Hoyland10, Ricardo Génova-Santos11.
Abstract
This paper presents preliminary polarization measurements and systematic-error characterization of the Thirty Gigahertz Instrument receiver developed for the QUIJOTE experiment. The instrument has been designed to measure the polarization of Cosmic Microwave Background radiation from the sky, obtaining the Q, U, and I Stokes parameters of the incoming signal simultaneously. Two kinds of linearly polarized input signals have been used as excitations in the polarimeter measurement tests in the laboratory; these show consistent results in terms of the Stokes parameters obtained. A measurement-based systematic-error characterization technique has been used in order to determine the possible sources of instrumental errors and to assist in the polarimeter calibration process.Entities:
Keywords: astronomy; cosmic microwave background; instrumentation; polarization; systematic errors
Year: 2015 PMID: 26251906 PMCID: PMC4570362 DOI: 10.3390/s150819124
Source DB: PubMed Journal: Sensors (Basel) ISSN: 1424-8220 Impact factor: 3.576
Figure 1TGI polarimeter block diagram.
Detected voltages for an x-axis linearly polarized signal.
| ΦT | Vd1 | Vd2 | Vd3 | Vd4 |
|---|---|---|---|---|
| 0º | K·(I + Q) | K·(I − Q) | K·(I + U) | K·(I − U) |
| 90º | K·(I + U) | K·(I − U) | K·(I − Q) | K·(I + Q) |
| 180º | K·(I − Q) | K·(I + Q) | K·(I − U) | K·(I + U) |
| 270º | K·(I − U) | K·(I + U) | K·(I + Q) | K·(I − Q) |
Figure 2TGI Polarimeter measurement test bench. (a) Receiver functionality test bench (BEM details in Figure 1). (b) Polarimeter test bench with x-axis linearly polarized source. (c) y-axis polarized source.
Detected voltage values for each of the phase-states forming a saw-tooth waveform.
| State | vd (V) | |||
|---|---|---|---|---|
| 0 | 0 | 0 | 0 | 5.28 |
| 1 | 90 | 90 | 0 | 2.71 |
| 2 | 180 | 180 | 0 | 0.19 |
| 3 | 270 | 270 | 0 | 2.54 |
| 4 | 0 | 90 | 90 | 4.66 |
| 5 | 90 | 180 | 90 | 2.68 |
| 6 | 180 | 270 | 90 | 0.17 |
| 7 | 270 | 0 | 90 | 2.54 |
| 8 | 0 | 180 | 180 | 5.26 |
| 9 | 90 | 270 | 180 | 2.62 |
| 10 | 180 | 0 | 180 | 0.18 |
| 11 | 270 | 90 | 180 | 2.47 |
| 12 | 0 | 270 | 270 | 4.67 |
| 13 | 90 | 0 | 270 | 2.63 |
| 14 | 180 | 90 | 270 | 0.17 |
| 15 | 270 | 180 | 270 | 2.66 |
Figure 3Detected voltage waveform from Table 2 values (black trace) and the resulting waveform calculated from the FFT of the measured values (blue trace).
Figure 4Detected signal before (a) and after (b) phase-error correction.
Figure 5First polarimeter example: Detected signals resulting from the phase-adjustment process with an x-axis polarized input signal (a) and measurement process with a y-axis polarized input signal (b).
Parameters calculated from the detected signals shown in Figure 5 (Phase in Deg. and ISO in dB).
| Detector | Ux | Qx | Phasex | ISOx | Ix | Uy | Qy | Phasey | ISOy | Iy |
|---|---|---|---|---|---|---|---|---|---|---|
| 1 | −2.1·10−2 | 0.92 | −1.295 | −16.46 | 0.92 | 2.2·10−2 | −0.93 | 178.64 | −16.25 | 0.93 |
| 2 | −1.0·10−2 | 0.89 | 179.34 | −19.41 | 0.89 | 1.2·10−2 | −0.89 | −0.79 | −18.61 | 0.89 |
| 3 | 0.13 | 0.93 | −82.19 | −8.63 | 0.94 | −0.13 | −0.93 | 97.79 | −8.64 | 0.94 |
| 4 | 0.11 | 0.94 | 96.78 | −9.21 | 0.94 | −0.11 | −0.94 | −83.18 | −9.22 | 0.95 |
Parameters calculated by using only the states numbered from 12 to 15 in Table 2 (Phase in Deg. and ISO in dB).
| Detector | Ux | Qx | Phasex | ISOx | Ix | Uy | Qy | Phasey | ISOy | Iy |
|---|---|---|---|---|---|---|---|---|---|---|
| 1 | 5.72·10−3 | 0.89 | 0.37 | −21.91 | 0.89 | −1.24·10−3 | −0.9 | −179.92 | −28.62 | 0.9 |
| 2 | 1.07·10−2 | 0.87 | −179.29 | −19.09 | 0.87 | −1.62·10−2 | −0.86 | 1.08 | −17.26 | 0.86 |
| 3 | 1.13·10−2 | 0.94 | −89.31 | −19.22 | 0.94 | −2.87·10−3 | −0.97 | 90.17 | −25.27 | 0.97 |
| 4 | −1.22·10−2 | 0.98 | 89.27 | −18.95 | 0.98 | −6.6·10−3 | −0.95 | −90.41 | −21.48 | 0.95 |
Figure 6Second polarimeter example: Detected signals resulting from the phase-adjustment (a) and measurement (b) processes.
Parameters calculated from the detected signals shown in Figure 6 (Phase in Deg. and ISO in dB).
| Detector | Ux | Qx | Phasex | ISOx | Ix | Uy | Qy | Phasey | ISOy | Iy |
|---|---|---|---|---|---|---|---|---|---|---|
| 1 | 3.71·10−3 | 0.94 | 0.23 | −23.9 | 0.94 | 6.73·10−3 | −0.92 | 179.58 | −21.33 | 0.92 |
| 2 | 4.61·10−3 | 0.95 | −179.73 | −23.25 | 0.95 | 4.43·10−2 | −0.93 | −0.27 | −23.19 | 0.93 |
| 3 | −1.08·10−3 | 0.96 | −90.07 | −29.18 | 0.96 | 8.84·10−3 | −0.94 | 89.47 | −20.35 | 0.94 |
| 4 | 2.8·10−2 | 0.96 | 91.67 | −15.35 | 0.96 | −2·10−2 | −0.94 | −88.87 | −16.68 | 0.94 |
Figure 7Simplified diagram of the polarimeter parametric model.
Ideal parametric model simulation results using four excitation signals with polarization angles of 0°, 45°, 90°, and 135°.
| 0° | 100.00 | −60.7 | 5·10−5 | 100.00 | −60.6 | −180 |
| 45° | 100.00 | −58.4 | −90 | 100.00 | −58.4 | 90 |
| 90° | 100.00 | −60.7 | −180 | 100.00 | −60.7 | 5·10−5 |
| 135° | 100.00 | −65.8 | 90 | 100.00 | −65.8 | −90 |
| 0° | 100.00 | −122.2 | −90 | 100.00 | −122.3 | 90 |
| 45° | 100.00 | −62.2 | −180 | 100.00 | −62.2 | 3·10−5 |
| 90° | 100.00 | −121.4 | 90 | 100.00 | −122.4 | −90 |
| 135° | 100.00 | −62.2 | −3·10−5 | 100.00 | −62.2 | 180 |
Figure 8Optimized parametric model simulation results. (a) Fit of the situation in Figure 5a; (b) Fit of the situation in Figure 6a.
Model parameters fitting the measurement results in Figure 5a.
| Parameter | Value ( | Value ( | |
|---|---|---|---|
| Att (dB) | 1.80 | 0.75 | |
| Att2 (dB) | 1.00 | 0.90 | |
| GB (dB) | −0.54 | 0.29 | |
| GB2 (dB) | −0.19 | 0.32 | |
| Phs (°) | 0.98 | −0.25 | |
| Phs2 (°) | −8.27 | −0.55 | |
| PB (°) | −0.37 | −0.07 | |
| PB2 (°) | −0.49 | 0.89 |