| Literature DB >> 26248272 |
Kristian Rink1, Moritz C Berger2, Andreas Korzowski2, Mathies Breithaupt2, Armin Biller3, Peter Bachert2, Armin M Nagel2.
Abstract
The purpose of this study is to develop nuclear-Overhauser-enhanced (NOE) [(1)H]-(31)P magnetic resonance imaging (MRI) based on 3D fully-balanced steady-state free precession (fbSSFP). Therefore, two implementations of a 3D fbSSFP sequence are compared using frequency-selective excitation (FreqSel) and multipoint-Dixon (MP-Dixon). (31)P-containing model solutions and four healthy volunteers were examined at field strengths of B0=3T and 7T. Maps of the distribution of phosphocreatine (PCr), inorganic phosphate (Pi), and adenosine 5´-triphosphate (ATP) in the human calf were obtained with an isotropic resolution of 1.5cm (1.0cm) in an acquisition time of 5min (10min). NOE-pulses had the highest impact on the PCr acquisitions enhancing the signal up to (82 ± 13) % at 3T and up to (37 ± 9) % at 7T. An estimation of the level of PCr in muscle tissue from [(1)H]-(31)P MRI data yielded a mean value of (33 ± 8) mM. In conclusion, direct [(1)H]-(31)P imaging using FreqSel as well as MP-Dixon is possible in clinically feasible acquisition times. FreqSel should be preferred for measurements where only a single metabolite resonance is considered. MP-Dixon performs better in terms of SNR if a larger spectral width is of interest.Entities:
Keywords: (31)P MRI; Dixon; High-energy phosphate metabolism; Nuclear Overhauser effect; Steady-state free precession (SSFP)
Mesh:
Substances:
Year: 2015 PMID: 26248272 DOI: 10.1016/j.mri.2015.07.017
Source DB: PubMed Journal: Magn Reson Imaging ISSN: 0730-725X Impact factor: 2.546