Overcoming the problem of residual microbial contamination in dental suction units left by conventional disinfection using novel single component suction handpieces in combination with automated flood disinfection.

M A Boyle1, M J O'Donnell1, R J Russell2, N Galvin3, J Swan1, D C Coleman4.   

Abstract

OBJECTIVES: Decontaminating dental chair unit (DCU) suction systems in a convenient, safe and effective manner is problematic. This study aimed to identify and quantify the extent of the problems using 25 DCUs, methodically eliminate these problems and develop an efficient approach for reliable, effective, automated disinfection.
METHODS: DCU suction system residual contamination by environmental and human-derived bacteria was evaluated by microbiological culture following standard aspiration disinfection with a quaternary ammonium disinfectant or alternatively, a novel flooding approach to disinfection. Disinfection of multicomponent suction handpieces, assembled and disassembled, was also studied. A prototype manual and a novel automated Suction Tube Cleaning System (STCS) were developed and tested, as were novel single component suction handpieces.
RESULTS: Standard aspiration disinfection consistently failed to decontaminate DCU suction systems effectively. Semi-confluent bacterial growth (101-500 colony forming units (CFU) per culture plate) was recovered from up to 60% of suction filter housings and from up to 19% of high and 37% of low volume suction hoses. Manual and automated flood disinfection of DCU suction systems reduced this dramatically (ranges for filter cage and high and low volume hoses of 0-22, 0-16 and 0-14CFU/plate, respectively) (P<0.0001). Multicomponent suction handpieces could not be adequately disinfected without prior removal and disassembly. Novel single component handpieces, allowed their effective disinfection in situ using the STCS, which virtually eliminated contamination from the entire suction system.
CONCLUSION: Flood disinfection of DCU suction systems and single component handpieces radically improves disinfection efficacy and considerably reduces potential cross-infection and cross-contamination risks. CLINICAL SIGNIFICANCE: DCU suction systems become heavily contaminated during use. Conventional disinfection does not adequately control this. Furthermore, multicomponent suction handpieces cannot be adequately disinfected without disassembly, which is costly in time, staff and resources. The automated STCS DCU suction disinfection system used with single component handpieces provides an effective solution.
Copyright © 2015 The Authors. Published by Elsevier Ltd.. All rights reserved.

Entities:  

Keywords:  Automated disinfection; Cross-contamination; Decontamination; Dental suction handpieces; Dental unit suction; Disinfection; Infection prevention; Infection reservoir; Risk reduction

Mesh:

Substances:

Year:  2015        PMID: 26248229     DOI: 10.1016/j.jdent.2015.07.018

Source DB:  PubMed          Journal:  J Dent        ISSN: 0300-5712            Impact factor:   4.379


  2 in total

1.  Portable Medical Suction and Aspirator Devices: Are the Design and Performance Standards Relevant?

Authors:  Saketh R Peri; Forhad Akhter; Robert A De Lorenzo; R Lyle Hood
Journal:  Sensors (Basel)       Date:  2022-03-25       Impact factor: 3.576

2.  Suction Hoses of Dental Units as a Potential Source of Microbial Contamination.

Authors:  Tayebeh Zeinali; Elham Bozorgvar; Moghgan Habibi; Narjes Akbari; Behnam Barikbin
Journal:  Oman Med J       Date:  2020-03-16
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.