Literature DB >> 26233532

Preliminary testing of a novel bilateral plating technique for treating periprosthetic fractures of the distal femur.

Aaron Muizelaar1, Mitchell J Winemaker2, Cheryl E Quenneville3, Gregory R Wohl4.   

Abstract

BACKGROUND: Current stabilization methods for periprosthetic fractures of the distal femur are inadequate in achieving fracture fixation, with complication rates as high as 29%. A major contributor to poor outcomes is that these methods rely only on screw purchase in the bone to maintain fracture reduction. We designed, manufactured and evaluated a novel plating method that utilizes the femoral prosthesis to enhance stability for treatment of distal femoral periprosthetic fractures.
METHODS: Medial and lateral plates were designed and manufactured based on geometry of a synthetic femur and femoral prosthesis. The two plates were linked via a compression screw and a small tab on each plate that inserted into pre-existing slots on the prosthesis. Mechanical tests (500N compression or 250N anterior directed cantilever bending), were performed on synthetic femurs with simple transverse fractures (3mm gap) just superior to the distal femoral prosthesis that were stabilized using either the prototype plates or a single lateral plate. Translational movements of the fracture site during loading were measured using 3D motion tracking.
FINDINGS: With the single lateral plate, the distal fragment experienced a resultant displacement of 0.40mm under cantilever bending and 0.61mm under compression (13% and 20% respectively of fracture gap width). With the bilateral plates, fracture gap motion was significantly reduced to 0.13mm under bending and compression (4.3% of the fracture gap).
INTERPRETATION: Our results indicate that a bilateral plating method is capable of improving stabilization of periprosthetic fractures compared to the traditional lateral plating technique.
Copyright © 2015 Elsevier Ltd. All rights reserved.

Entities:  

Keywords:  Bending; Bilateral plates; Compression; Distal femur; Fracture displacement; Lateral locking plate; Periprosthetic fracture

Mesh:

Year:  2015        PMID: 26233532     DOI: 10.1016/j.clinbiomech.2015.07.008

Source DB:  PubMed          Journal:  Clin Biomech (Bristol, Avon)        ISSN: 0268-0033            Impact factor:   2.063


  3 in total

1.  A case series of mortality and morbidity in distal femoral periprosthetic fractures.

Authors:  Toby Jennison; Rathan Yarlagadda
Journal:  J Orthop       Date:  2019-09-11

Review 2.  Double fixation for complex distal femoral fractures.

Authors:  Karl Stoffel; Christoph Sommer; Mark Lee; Tracy Y Zhu; Karsten Schwieger; Christopher Finkemeier
Journal:  EFORT Open Rev       Date:  2022-04-21

3.  Analysis of risk factors for revision in distal femoral fractures treated with lateral locking plate: a retrospective study in Chinese patients.

Authors:  Guojin Hou; Fang Zhou; Yun Tian; Hongquan Ji; Zhishan Zhang; Yan Guo; Yang Lv; Zhongwei Yang; Yawen Zhang
Journal:  J Orthop Surg Res       Date:  2020-08-12       Impact factor: 2.359

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.