Literature DB >> 26233490

Evaluation of ultraviolet irradiation efficacy in an automated system for the aseptic compounding using challenge test.

Francesca Bruscolini1, Demis Paolucci2, Valeria Rosini2, Luigia Sabatini1, Elisa Andreozzi1, Anna Pianetti1.   

Abstract

OBJECTIVE: Ultraviolet (UV) irradiation efficacy in the intravenous compounding robot APOTECAchemo was evaluated to define the best operative conditions in terms of sterility and time optimization.
DESIGN: The challenge test was used against Pseudomonas aeruginosa, Escherichia coli, Staphylococcus aureus, Bacillus subtilis spores and Candida albicans. Inoculated plates were placed inside the robot and irradiated for different times. Microbial air and surface quality inside the equipment were monitored utilizing settle and contact plates, swabs.
RESULTS: After 4 h, no microorganisms were viable with killing rates ranging from 5- to 7-log for different microorganisms after 1 h of exposition. In confirmation of the efficacy of the UV irradiation program adopted, the microbial monitoring inside the equipment always gave negative results.
CONCLUSIONS: This is the first exhaustive investigation of UV irradiation efficacy in the aseptic pharmaceutical production. We demonstrated that UV irradiation plays an essential role in maintaining the sterility condition of the workplace inside the APOTECAchemo and assuring the standards for aseptic manufacturing of medicinal drugs, as required for Class A clean areas. A 4-h UV irradiation also ensures sterility in the case of very resistant microorganisms and in the presence of high microbial charge (10(8) CFU/ml), but a killing rate of 5 or more is already recorded after the first hour of exposition. The results provide useful information for the best operative conditions in terms of both sterility and time optimization, not only for the automated compounding, but also for the traditional aseptic manufacturing processes.
© The Author 2015. Published by Oxford University Press in association with the International Society for Quality in Health Care; all rights reserved.

Entities:  

Keywords:  UV efficiency; aseptic drug compounding; challenge test

Mesh:

Year:  2015        PMID: 26233490     DOI: 10.1093/intqhc/mzv051

Source DB:  PubMed          Journal:  Int J Qual Health Care        ISSN: 1353-4505            Impact factor:   2.038


  2 in total

1.  Microbiological validation of a robot for the sterile compounding of injectable non-hazardous medications in a hospital environment.

Authors:  Luigia Sabatini; Demis Paolucci; Francesco Marinelli; Anna Pianetti; Monica Sbaffo; Celestino Bufarini; Maurizio Sisti
Journal:  Eur J Hosp Pharm       Date:  2019-02-04

Review 2.  Review of Decontamination Techniques for the Inactivation of Bacillus anthracis and Other Spore-Forming Bacteria Associated with Building or Outdoor Materials.

Authors:  Joseph P Wood; Alden Charles Adrion
Journal:  Environ Sci Technol       Date:  2019-04-02       Impact factor: 9.028

  2 in total