A multicenter study to quantify systematic variations and associated uncertainties in source positioning with commonly used HDR afterloaders and ring applicators for the treatment of cervical carcinomas.

O Awunor1, D Berger2, C Kirisits3.   

Abstract

PURPOSE: The reconstruction of radiation source position in the treatment planning system is a key part of the applicator reconstruction process in high dose rate (HDR) brachytherapy treatment of cervical carcinomas. The steep dose gradients, of as much as 12%/mm, associated with typical cervix treatments emphasize the importance of accurate and precise determination of source positions. However, a variety of methodologies with a range in associated measurement uncertainties, of up to ±2.5 mm, are currently employed by various centers to do this. In addition, a recent pilot study by Awunor et al. ["Direct reconstruction and associated uncertainties of (192)Ir source dwell positions in ring applicators using gafchromic film in the treatment planning of HDR brachytherapy cervix patients," Phys. Med. Biol. 58, 3207-3225 (2013)] reported source positional differences of up to 2.6 mm between ring sets of the same type and geometry. This suggests a need for a comprehensive study to assess and quantify systematic source position variations between commonly used ring applicators and HDR afterloaders across multiple centers.
METHODS: Eighty-six rings from 20 European brachytherapy centers were audited in the form of a postal audit with each center collecting the data independently. The data were collected by setting up the rings using a bespoke jig and irradiating gafchromic films at predetermined dwell positions using four afterloader types, MicroSelectron, Flexitron, GammaMed, and MultiSource, from three manufacturers, Nucletron, Varian, and Eckert & Ziegler BEBIG. Five different ring types in six sizes (Ø25-Ø35 mm) and two angles (45° and 60°) were used. Coordinates of irradiated positions relative to the ring center were determined and collated, and source position differences quantified by ring type, size, and angle.
RESULTS: The mean expanded measurement uncertainty (k = 2) along the direction of source travel was ±1.4 mm. The standard deviation associated with the source position reproducibility was within ±1.0 mm for all afterloaders. Maximum source positional variations of 2.1 and 3.9, 1.8 and 5.4, and 2.3 and 3.4 mm were observed at standard treatment positions for the Ø26, Ø30, and Ø32 mm sized 45° and 60° rings, respectively. Mean positional differences between a majority of the rings were within ±1.0 mm. Mean positional differences between a majority of the intracenter ring sets were within the expanded measurement uncertainty. When comparing the 45°-60° source paths, mean differences of 1.6, 0.9, and 0.9 mm were observed across the Ø26, Ø30 (MicroSelectron), and Ø32 mm (GammaMed) rings, respectively. When comparing to manufacturer source path models, maximum offsets of 1.9 and 2.1, 2.6 and 2.3, and 0.8 and 1.6 mm were observed for the Ø26, Ø30 (MicroSelectron), and Ø30 mm (Flexitron) sized 45° and 60° rings, respectively. When comparing the audit to ring commissioning data of participating centers, mean differences of up to 2.4 mm were observed.
CONCLUSIONS: A majority of the audited rings showed a good degree of manufacturer consistency; however, substantial positional variation observed between some rings emphasizes the importance of commissioning each ring before clinical use. Differences observed between audit and commissioning data also indicate some variation in source treatment positions across centers.

Entities:  

Mesh:

Year:  2015        PMID: 26233177     DOI: 10.1118/1.4923173

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  4 in total

Review 1.  Review of strategies for MRI based reconstruction of endocavitary and interstitial applicators in brachytherapy of cervical cancer.

Authors:  José Richart; Vicente Carmona-Meseguer; Teresa García-Martínez; Antonio Herreros; Antonio Otal; Santiago Pellejero; Ana Tornero-López; José Pérez-Calatayud
Journal:  Rep Pract Oncol Radiother       Date:  2018-07-23

2.  Impact of different Ir-192 source models on dose calculations in high-dose-rate brachytherapy.

Authors:  Yuki Otani; Shigetoshi Shimamoto; Iori Sumida; Yutaka Takahashi; Shoji Tani; Tetsuya Oshima; Satoshi Onosaka; Fumiaki Isohashi; Keisuke Tamari; Kazuhiko Ogawa
Journal:  Phys Imaging Radiat Oncol       Date:  2018-08-31

3.  Independent assessment of source position for gynecological applicator in high-dose-rate brachytherapy.

Authors:  Hiroyuki Okamoto; Satoshi Nakamura; Shie Nishioka; Kotaro Iijima; Akihisa Wakita; Yukinao Abe; Naoki Tohyama; Shinji Kawamura; Toshiyuki Minemura; Jun Itami
Journal:  J Contemp Brachytherapy       Date:  2017-10-19

4.  Dosimetric effects of the Smit sleeve on high-dose-rate brachytherapy tandem and ovoids plans for patients with locally advanced cervical cancer.

Authors:  Shahil Mehta; Benjamin Farnia; Alberto de la Zerda; Robabeh Rahimi; Aaron Wolfson; Lorraine Portelance
Journal:  J Contemp Brachytherapy       Date:  2019-12-06
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.