Therapeutic concentrations of hydroxocobalamin interferes with several spectrophotometric assays on the Beckman Coulter DxC and AU680 chemistry analyzers.

Pratistha Ranjitkar1, Dina N Greene1.   

Abstract

BACKGROUND: High doses of hydroxocobalamin (OHCob) are used to treat cyanide poisoning and cardiac complications. Since OHCob absorbs at multiple wavelengths often used in colorimetric assays, spurious laboratory results are likely to occur. The objective of this study was to examine interference caused by OHCob in colorimetric assays measured using the Beckman Coulter DxC and AU680.
METHODS: OHCob was dissolved in water and spiked into pooled "healthy" and "unhealthy" patient samples at two different concentrations (0.15 and 1.5 mg/mL). Spiked and unspiked samples were analyzed on both instruments and bias was calculated. A total of 23 analytes were tested on the DxC and 27 analytes were tested on the AU680. For analytes showing a bias ≥ 10%, OHCob was titrated from 0.2-1.5mg/mL.
RESULTS: The following analytes were affected on the DxC and AU680: alanine aminotransferase, amylase, total bilirubin, cholesterol, creatine kinase, creatinine, magnesium, uric acid. Direct bilirubin, iron, phosphate, total protein and triglycerides were only affected on the DxC. Biases observed were positive or negative and fixed or proportional.
CONCLUSIONS: Between the DxC and AU680, several analytes were affected at therapeutic OHCob concentrations. Hence, it is important for laboratories to know how their instruments are affected, and for clinicians to alert the lab when these samples are expected.
Copyright © 2015. Published by Elsevier B.V.

Entities:  

Keywords:  AU680; Beckman Coulter; Colorimetric assays; DxC; Hydroxocobalamin; Interference

Mesh:

Substances:

Year:  2015        PMID: 26232160     DOI: 10.1016/j.cca.2015.07.024

Source DB:  PubMed          Journal:  Clin Chim Acta        ISSN: 0009-8981            Impact factor:   3.786


  1 in total

1.  Hemolysis index to detect degree of hydroxocobalamin interference with common laboratory tests.

Authors:  Laura Fueyo; Juan Robles; Irene Aguilar; Aina M Yáñez; Magdalena Socias; Magdalena Parera
Journal:  J Clin Lab Anal       Date:  2016-11-10       Impact factor: 2.352

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.