Corneal Power Measurement Obtained by Fourier-Domain Optical Coherence Tomography: Repeatability, Reproducibility, and Comparison With Scheimpflug and Automated Keratometry Measurements.

Qinmei Wang1, Yanjun Hua, Giacomo Savini, Hao Chen, Fangjun Bao, Shishi Lin, Weicong Lu, Jinhai Huang.   

Abstract

PURPOSE: To assess the repeatability and reproducibility of corneal power values obtained by a Fourier-domain optical coherence tomography (FD-OCT) system (RTVue) and to compare them with the values obtained by a Scheimpflug camera system (Pentacam HR) and by automated keratometry (IOL Master).
METHODS: Thirty-two eyes from 32 healthy subjects were included in this prospective study. Two experienced observers measured each eye 3 consecutive times with the Pentacam, IOLMaster, and RTVue centered on either the pupil or corneal vertex. The conventional keratometry equivalent (CKE) and anterior (Ka), posterior (Kp), and net (Kn) corneal power values were determined.
RESULTS: The corneal power values obtained by the RTVue showed high repeatability (all intraclass correlation coefficient >0.96) and reproducibility (coefficient of variation <1.0%). Pupil-centered FD-OCT performed slightly better than corneal vertex-centered FD-OCT. Mean corneal values had higher reproducibly than any of the individual values. CKE, Ka, Kp, and Kn obtained by FD-OCT were 0.62 to 0.68 diopters (D), 0.70 to 0.76 D, 0.11 to 0.13 D, and 0.93 to 0.94 D higher than those obtained by the Pentacam HR, respectively. CKE and Ka obtained with the RTVue were also 0.60 to 0.74 D higher than those obtained with the IOLMaster, respectively.
CONCLUSIONS: The corneal power measurements obtained by the RTVue FD-OCT system showed high repeatability and reproducibility. Measurements obtained by FD-OCT with pupil centration were more reproducible than those obtained by FD-OCT with corneal vertex centration. We recommend that pupil-centered FD-OCT be used in clinical applications. Neither RTVue versus Pentacam HR nor RTVue versus IOLMaster can be used interchangeably.

Entities:  

Mesh:

Year:  2015        PMID: 26226469     DOI: 10.1097/ICO.0000000000000564

Source DB:  PubMed          Journal:  Cornea        ISSN: 0277-3740            Impact factor:   2.651


  3 in total

Review 1.  Applications of Anterior Segment Optical Coherence Tomography in Cornea and Ocular Surface Diseases.

Authors:  Sang Beom Han; Yu-Chi Liu; Karim Mohamed Noriega; Jodhbir S Mehta
Journal:  J Ophthalmol       Date:  2016-09-19       Impact factor: 1.909

2.  Precision of a new ocular biometer in eyes with cataract using swept source optical coherence tomography combined with Placido-disk corneal topography.

Authors:  Wenwen Wang; Yaxin Miao; Giacomo Savini; Colm McAlinden; Hao Chen; Qingjie Hu; Qinmei Wang; Jinhai Huang
Journal:  Sci Rep       Date:  2017-10-23       Impact factor: 4.379

3.  The precision and agreement of corneal thickness and keratometry measurements with SS-OCT versus Scheimpflug imaging.

Authors:  Yune Zhao; Ding Chen; Giacomo Savini; Qing Wang; Hongfang Zhang; Yili Jin; Benhao Song; Rui Ning; Jinhai Huang; Chenyang Mei
Journal:  Eye Vis (Lond)       Date:  2020-06-09
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.