Accuracy of a Simplified Glucose Measurement Device--The HemoCue Glucose 201RT.

Ellen Segerhag1, Viveca Gyberg2, Kyriacos Ioannides3, Catriona Jennings3, Lars Rydén2, Martin Stagmo4, Camilla Hage1,2.   

Abstract

BACKGROUND: Easily available, accurate glucose recordings are important when screening for and managing people with diabetes. The photometric HemoCue(®) (Ängelholm, Sweden) Glucose 201+ system, which delivers lab-comparable glucose recordings, has the drawback that its microcuvettes must be delivered and stored at 4-8°C. A newly developed system, HemoCue Glucose 201RT, has microcuvettes that can be stored at room temperature. SUBJECTS AND METHODS: Participants (n=444; 18-80 years old) in the EUROASPIRE IV survey, all with coronary artery disease, some with known diabetes, were investigated. Plasma glucose recordings, fasting in all participants and postprandial in the majority, were simultaneously recorded with both pieces of equipment. Congruence was expressed as median absolute difference and median absolute relative difference between the two sets of equipment and also compared according to the International Organization for Standardization (ISO) 15197:2013 criteria. Clinical accuracy was calculated with Clarke error grid analysis and cross-tabulated while considering different glucose categories (normal, impaired glucose tolerance, and diabetes).
RESULTS: The median absolute difference between the two devices was +0.1 mmol/L, and the median absolute relative difference was +5.4%. This also corresponded with the ISO criteria. In the Clarke error grid, 99.8% ended up in Zones A and B, and 90% of the glucose values in the cross-table allocated the participant to the same glucose category.
CONCLUSIONS: The HemoCue Glucose 201RT system is accurate, with small nonsystematic deviations, when compared with the commonly used HemoCue Glucose 201+. It is predicted that the HemoCue Glucose 201RT, which is more user friendly, will be a preferred alternative to the HemoCue Glucose 201+.

Entities:  

Mesh:

Substances:

Year:  2015        PMID: 26218499     DOI: 10.1089/dia.2014.0354

Source DB:  PubMed          Journal:  Diabetes Technol Ther        ISSN: 1520-9156            Impact factor:   6.118


  3 in total

1.  Mannose-binding lectin does not explain the dismal prognosis after an acute coronary event in dysglycaemic patients. A report from the GAMI cohort.

Authors:  Sara Meziani; Giulia Ferrannini; Mette Bjerre; Troels K Hansen; Viveca Ritsinger; Anna Norhammar; Viveca Gyberg; Per Näsman; Lars Rydén; Linda G Mellbin
Journal:  Cardiovasc Diabetol       Date:  2022-07-08       Impact factor: 8.949

2.  Diagnosis of Gestational Diabetes Mellitus with Point-of-Care Methods for Glucose versus Hospital Laboratory Method Using Isotope Dilution Gas Chromatography-Mass Spectrometry as Reference.

Authors:  Karl Kristensen; Anne-Marie Wangel; Anastasia Katsarou; Nael Shaat; David Simmons; Helena Fadl; Kerstin Berntorp
Journal:  J Diabetes Res       Date:  2020-03-20       Impact factor: 4.011

3.  Standing up against office sitting: A study protocol.

Authors:  Philippe Gradidge; Merling Phaswana; Katrien Wijndaele; Nigel Crowther; Catherine Draper
Journal:  S Afr J Physiother       Date:  2020-09-04
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.