Literature DB >> 26218183

Predictive factors of lead failure in patients implanted with cardiac devices.

Yoshiyasu Aizawa1, Masachika Negishi2, Shin Kashimura2, Kazuaki Nakajima2, Akira Kunitomi2, Yoshinori Katsumata2, Takahiko Nishiyama2, Takehiro Kimura2, Nobuhiro Nishiyama2, Kotaro Fukumoto2, Yoko Tanimoto2, Shun Kohsaka2, Seiji Takatsuki2, Keiichi Fukuda2.   

Abstract

INTRODUCTION: Lead failures (LFs) are one of the most common complications in patients implanted with cardiovascular implantable electronic devices. LFs often cause serious secondary complications such as inappropriate ICD shocks or asystole. This study aimed to identify the clinical factors associated with the occurrence of LFs.
METHODS: A total of 735 consecutive device implantations (mean age 67±15years, males 64%) performed at a single university hospital setting from 1997 to 2014 were included. The implanted devices consisted of 421 pacemakers, 250 implantable cardioverter defibrillators (ICD), 9 cardiac resynchronization therapy pacemakers (CRT-P), and 55 CRT defibrillators (CRT-D). The primary endpoint was the development of an LF.
RESULTS: During a mean duration of 5.8±4.3years, 38 LFs developed in 31 patients (mean age 56±14years). LFs included 32 ICD (7 Sprint Fidelis, 2 Riata), and 6 pacing leads. Nine patients received inappropriate ICD shocks and 1 had syncope due to an LF. All patients underwent lead reinsertions with device replacements. Eight patients required opposite site implantations due to venous occlusions. The predictive factors of LFs were the age, male sex, taller body length, ICD vs. pacemaker, lesser lead number, extra-thoracic puncture of the axillary vein vs. a cut-down of the cephalic vein, use of recalled leads and patients with idiopathic ventricular fibrillation (IVF) and Brugada syndrome (BrS).
CONCLUSION: LFs occurred mainly with ICD leads. A lesser age, the puncture method, lead model, and diagnosis of IVF/BrS were associated with the development of LFs.
Copyright © 2015 Elsevier Ireland Ltd. All rights reserved.

Entities:  

Keywords:  Brugada syndrome; Fidelis; Idiopathic ventricular fibrillation; Implantable cardioverter defibrillator; Lead failure; Pacemaker; Riata

Mesh:

Year:  2015        PMID: 26218183     DOI: 10.1016/j.ijcard.2015.07.055

Source DB:  PubMed          Journal:  Int J Cardiol        ISSN: 0167-5273            Impact factor:   4.164


  6 in total

1.  Efficacy of ultrasound-guided axillary/subclavian venous approaches for pacemaker and defibrillator lead implantation: a randomized study.

Authors:  Mattia Liccardo; Pasquale Nocerino; Salzano Gaia; Carmine Ciardiello
Journal:  J Interv Card Electrophysiol       Date:  2018-01-15       Impact factor: 1.900

2.  Optimized Axillary Vein Technique versus Subclavian Vein Technique in Cardiovascular Implantable Electronic Device Implantation: A Randomized Controlled Study.

Authors:  Peng Liu; Yi-Feng Zhou; Peng Yang; Yan-Sha Gao; Gui-Ru Zhao; Shi-Yan Ren; Xian-Lun Li
Journal:  Chin Med J (Engl)       Date:  2016-11-20       Impact factor: 2.628

3.  Long-term reliability of the defibrillator lead inserted by the extrathoracic subclavian puncture.

Authors:  Masaya Watanabe; Hisashi Yokoshiki; Hirofumi Mitsuyama; Kazuya Mizukami; Taro Tenma; Rui Kamada; Masayuki Takahashi; Yoshiro Matsui; Toshihisa Anzai
Journal:  J Arrhythm       Date:  2018-08-18

4.  The axillary vein and its tributaries are not in the mirror image of the axillary artery and its branches.

Authors:  HyeYeon Lee; JongHo Bang; SooJung Kim; HeeJun Yang
Journal:  PLoS One       Date:  2019-01-10       Impact factor: 3.240

5.  Effectiveness and safety of transvenous extraction of single- versus dual-coil implantable cardioverter-defibrillator leads at single-center experience.

Authors:  Andrzej Ząbek; Krzysztof Boczar; Maciej Dębski; Mateusz Ulman; Roman Pfitzner; Robert Musiał; Jacek Lelakowski; Barbara Małecka
Journal:  Medicine (Baltimore)       Date:  2019-07       Impact factor: 1.817

Review 6.  Complications of electrotherapy - the dark side of treatment with cardiac implantable electronic devices.

Authors:  Szymon Domagała; Michał Domagała; Jakub Chyła; Celina Wojciechowska; Marianna Janion; Anna Polewczyk
Journal:  Postepy Kardiol Interwencyjnej       Date:  2018-03-22       Impact factor: 1.426

  6 in total