Literature DB >> 26216229

A Novel Method for Assessment of Polyethylene Liner Wear in Radiopaque Tantalum Acetabular Cups: Clinical Validation in Patients Enrolled in a Randomized Controlled Trial.

Anders Troelsen1, Meridith E Greene2, David C Ayers3, Charles R Bragdon2, Henrik Malchau2.   

Abstract

Conventional radiostereometric analysis (RSA) for wear is not possible in patients with tantalum cups. We propose a novel method for wear analysis in tantalum cups. Wear was assessed by gold standard RSA and the novel method in total hip arthroplasty patients enrolled in a randomized controlled trial receiving either titanium or tantalum cups (n=46). The novel method estimated the center of the head using a model based on identification of two proximal markers on the stem and knowledge of the stem/head configuration. The novel method was able to demonstrate a pattern of wear that was similar to the gold standard in titanium cups. The novel method offered accurate assessment and is a viable solution for assessment of wear in studies with tantalum cups.
Copyright © 2015 Elsevier Inc. All rights reserved.

Entities:  

Keywords:  conventional polyethylene; highly cross-linked polyethylene; novel method for wear measurement; polyethylene wear; radiostereometric analysis (RSA); total hip replacement

Mesh:

Substances:

Year:  2015        PMID: 26216229     DOI: 10.1016/j.arth.2015.06.050

Source DB:  PubMed          Journal:  J Arthroplasty        ISSN: 0883-5403            Impact factor:   4.757


  1 in total

1.  A simple method of measuring the wear of explanted acetabular component inserts.

Authors:  L Krakow; A Klockow; E Roehner; S Brodt; H Eijer; J Bossert; G Matziolis
Journal:  Bone Joint Res       Date:  2017-09       Impact factor: 5.853

  1 in total