In vitro standardization of two different removal devices in cemented implant prosthesis.

Gianmario Schierano1, Carlo Manzella1, Giulio Menicucci1, Alessio Parrotta1, Elisabetta M Zanetti2, Alberto L Audenino3.   

Abstract

OBJECTIVES: The purpose of this study was to describe the force generated by two different removal devices used to retrieve cemented crowns on implant abutments. The influence of six different operators was evaluated.
MATERIAL AND METHODS: Three replicated Coronaflex(®) (Kaltenbach & Voigt GmbH, KaVo Dental GmbH) and reverse hammer setups were tested. The experimental setup has employed a screw bearing a diametral hole through which a loop holder passed. The screw was attached to a force transducer (Brüel & Kjaer, type 8201), and the loop holder arm was kept perpendicular to the transducer axis. The results were statistically evaluated with ANOVA.
RESULTS: The operator has resulted to play significant influence with reference to reverse hammer (coefficient of variation 43.3%) rather than with Coronaflex(®) (9.8%). Evaluating every single operator, more variation can still be found by considering each reverse hammer (37.5%) rather than each Coronaflex(®) (8.8%).
CONCLUSION: Coronaflex(®) device was found to systematically reach a more repeatable and higher peak amplitude of forces compared with reverse hammer, both by experienced and inexperienced operators.
© 2015 John Wiley & Sons A/S. Published by John Wiley & Sons Ltd.

Keywords:  Coronaflex®; FDP; force transducer; implant prosthesis; reverse hammer

Mesh:

Year:  2015        PMID: 26212204     DOI: 10.1111/clr.12671

Source DB:  PubMed          Journal:  Clin Oral Implants Res        ISSN: 0905-7161            Impact factor:   5.977


  7 in total

1.  Reliability, Learnability and Efficiency of Two Tools for Cement Crowns Retrieval in Dentistry.

Authors:  Cristina Bignardi; Elisabetta M Zanetti; Mara Terzini; Anna R Ciccola; Gianmario Schierano; Alberto L Audenino
Journal:  Open Biomed Eng J       Date:  2018-06-29

2.  Impact of cement type and abutment height on pull-off force of zirconia reinforced lithium silicate crowns on titanium implant stock abutments: an in vitro study.

Authors:  Lisa Müller; Angelika Rauch; Daniel R Reissmann; Oliver Schierz
Journal:  BMC Oral Health       Date:  2021-11-19       Impact factor: 2.757

3.  A Comparison of Biomechanical Properties of Implant-Retained Overdenture Based on Precision Attachment Type.

Authors:  Małgorzata Idzior-Haufa; Agnieszka A Pilarska; Wiesław Hędzelek; Piotr Boniecki; Krzysztof Pilarski; Barbara Dorocka-Bobkowska
Journal:  Materials (Basel)       Date:  2021-05-17       Impact factor: 3.623

4.  The Influence of Custom-Milled Framework Design for an Implant-Supported Full-Arch Fixed Dental Prosthesis: 3D-FEA Sudy.

Authors:  João Paulo Mendes Tribst; Amanda Maria de Oliveira Dal Piva; Roberto Lo Giudice; Alexandre Luiz Souto Borges; Marco Antonio Bottino; Ettore Epifania; Pietro Ausiello
Journal:  Int J Environ Res Public Health       Date:  2020-06-05       Impact factor: 3.390

5.  Influence of Luting Materials on the Retention of Cemented Implant-Supported Crowns: An In Vitro Study.

Authors:  Ella A Naumova; Felix Roth; Berit Geis; Christine Baulig; Wolfgang H Arnold; Andree Piwowarczyk
Journal:  Materials (Basel)       Date:  2018-09-28       Impact factor: 3.623

6.  In Vitro Impact Testing to Simulate Implant-Supported Prosthesis Retrievability in Clinical Practice: Influence of Cement and Abutment Geometry.

Authors:  Andrea T Lugas; Mara Terzini; Elisabetta M Zanetti; Gianmario Schierano; Carlo Manzella; Domenico Baldi; Cristina Bignardi; Alberto L Audenino
Journal:  Materials (Basel)       Date:  2020-04-09       Impact factor: 3.623

Review 7.  Comprehensive Review on Current and Future Regulatory Requirements on Wearable Sensors in Preclinical and Clinical Testing.

Authors:  Alice Ravizza; Carmelo De Maria; Licia Di Pietro; Federico Sternini; Alberto L Audenino; Cristina Bignardi
Journal:  Front Bioeng Biotechnol       Date:  2019-11-08
  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.