Literature DB >> 26210096

Radiation dose and quickness of needle CT-interventions using a laser navigation system (LNS) compared with conventional method.

T Gruber-Rouh1, B Schulz2, K Eichler2, Nagy N N Naguib3, T J Vogl2, S Zangos2.   

Abstract

PURPOSE: The aim of this study was to analyse the radiation dose and quickness of needle interventions using a Laser Navigation System (LNS-group) compared with conventional method (control-group).
MATERIALS AND METHODS: In this prospective, randomized, comparative study 58 patients (19 females, 39 males; mean age, 62.9 years) were punctured either with LNS (n=29) or with conventional method with a skin mark of the puncture site (n=29). In the LNS method the puncture site was marked with laser without additional CT. Thoracic and abdominal intervention was performed in 30 and 28 patients, respectively. Radiation dose and time of the procedures were analysed. Statistical significance was calculated according to the Mann-Whitney-U-test.
RESULTS: Mean target access path in the patients of the LNS group was 6.0 cm (range, 3.0-10.1cm) and in the control group 6.0 cm (range, 1.0-10.3 cm). Time duration of complete intervention in the LNS group was 20:25 min (range, 07:00-34:00 min) and in the control group 28:00 min (range, 13:00-51:00 min). The dose-length-product (DLP) of intervention scan of the LNS group was 42.3 mGy cm (range, 10-125 mGy cm), and of the control group 59.7 mGy cm (range, 25-176.42 mGy cm).
CONCLUSION: Using the LNS for CT-guided interventions results in faster intervention time with a lower dose.
Copyright © 2015 Elsevier Ireland Ltd. All rights reserved.

Entities:  

Keywords:  CT-guided interventions; Laser navigation system; Quickness of intervention; Radiation dose

Mesh:

Year:  2015        PMID: 26210096     DOI: 10.1016/j.ejrad.2015.07.004

Source DB:  PubMed          Journal:  Eur J Radiol        ISSN: 0720-048X            Impact factor:   3.528


  2 in total

1.  Evaluation of a novel, patient-mounted system for CT-guided needle navigation-an ex vivo study.

Authors:  Anna Mokry; Florian Willmitzer; Rafael Hostettler; Henning Richter; Patrick Kircher; Sibylle Kneissl; Stephan Wetzel
Journal:  Neuroradiology       Date:  2018-12-01       Impact factor: 2.804

2.  Computed tomography-guided biopsies in children: accuracy, efficiency and dose usage.

Authors:  Tatjana Gruber-Rouh; Axel Thalhammer; Thomas Klingebiel; Nour-Eldin A Nour-Eldin; Thomas J Vogl; Katrin Eichler; Nagy Naguib; Martin Beeres
Journal:  Ital J Pediatr       Date:  2017-01-06       Impact factor: 2.638

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.