Literature DB >> 26207994

High definition urethral pressure profilometry: Evaluating a novel microtip catheter.

Mario Klünder1, Bastian Amend2, Martin Vaegler2, Alexandra Kelp2, Ronny Feuer3, Karl-Dietrich Sievert4, Arnulf Stenzl2, Oliver Sawodny3, Michael Ederer3.   

Abstract

INTRODUCTION: Urethral pressure profilometry (UPP) is used in the diagnosis of stress urinary incontinence (SUI). SUI is a significant medical, social, and economic problem, affecting about 12.5% of the population. A novel microtip catheter was developed for UPP featuring an inclination sensor and higher angular resolution compared to systems in clinical use today. Therewith, the location of each measured pressure sample can be determined and the spatial pressure distribution inside the urethra reconstructed. In order to assess the performance and plausibility of data from the microtip catheter, we compare it to data from a double balloon air charged system.
MATERIALS AND METHODS: Both catheters are used on sedated female minipigs. Data from the microtip catheter are processed through a signal reconstruction algorithm, plotted and compared against data from the air-charged catheter.
RESULTS: The microtip catheter delivers results in agreement with previous comparisons of microtip and air-charged systems. It additionally provides a new level of detail in the reconstructed UPPs which may lead to new insights into the sphincter mechanism of minipigs.
CONCLUSIONS: The ability of air-charged catheters to measure pressure circumferentially is widely considered a main advantage over microtip catheters. However, directional pressure readings can provide additional information on angular fluctuations in the urethral pressure distribution. It is shown that the novel microtip catheter in combination with a signal reconstruction algorithm delivers plausible data. It offers the opportunity to evaluate urethral structures, especially the sphincter, in context of the correct location within the anatomical location of the pelvic floor. Neurourol. Urodynam. 35:888-894, 2016.
© 2015 Wiley Periodicals, Inc. © 2015 Wiley Periodicals, Inc.

Entities:  

Keywords:  digital signal processing; urethral pressure; urinary catheters; urodynamics

Mesh:

Year:  2015        PMID: 26207994     DOI: 10.1002/nau.22835

Source DB:  PubMed          Journal:  Neurourol Urodyn        ISSN: 0733-2467            Impact factor:   2.696


  3 in total

1.  Establishing and monitoring of urethral sphincter deficiency in a large animal model.

Authors:  Alexandra Kelp; Anika Albrecht; Bastian Amend; Mario Klünder; Philipp Rapp; Oliver Sawodny; Arnulf Stenzl; Wilhelm K Aicher
Journal:  World J Urol       Date:  2017-09-22       Impact factor: 4.226

2.  A Method for Recording Urethral Pressure Profiles in Female Rats.

Authors:  Shengfei Xu; Xiaohui Li; Lei Xu; Biao Chen; Huibing Tan; Guanghui Du
Journal:  PLoS One       Date:  2015-10-26       Impact factor: 3.240

3.  Signal processing in urodynamics: towards high definition urethral pressure profilometry.

Authors:  Mario Klünder; Oliver Sawodny; Bastian Amend; Michael Ederer; Alexandra Kelp; Karl-Dietrich Sievert; Arnulf Stenzl; Ronny Feuer
Journal:  Biomed Eng Online       Date:  2016-03-22       Impact factor: 2.819