Automated control of inspired oxygen in ventilated preterm infants: crossover physiological study.

Mithilesh Lal1, Win Tin1,2, Sunil Sinha1,2.   

Abstract

AIM: To evaluate the efficacy of automated control of the fraction of inspired oxygen (FiO2 ) in comparison with manual FiO2 control in maintaining target pulse oxygen saturation (SpO2 ) range.
METHODS: Crossover physiological study involving preterm infants requiring mechanical ventilation and supplemental oxygen. Each infant was studied for two consecutive 12 hours in a random sequence. Outcome measures included the proportion of time with SpO2 within and outside the target range of 90-95%, extreme hypoxaemia (< 80%) and hyperoxaemia (≥ 98%).
RESULTS: Complete data set was available in 27 infants. The percentage of time (median, IQR) within the target range was higher during automated control 72.8 (58.8-82.6) compared to manual control 59.6 (49.3-73.3), p = 0.031. Corresponding reduction in per cent time below the target range was 18.1 (12.7-23.6) versus 25.9 (17.8-30.7), p = 0.028, and above the target range 4.8 (3-16) versus 10.1 (6.4-22.5), p = 0.026. Median (IQR) per cent time spent with severe hypoxaemia (SpO2 < 80%) and severe hyperoxaemia (SpO2 ≥ 98%) was 1.3 (0.1-2.9) versus 3.2 (1.4-6.1), p = 0.022 and 0.08 (0.05-0.36) versus 1.7 (0.7-6.8), p = 0.001, respectively. Median number of manual adjustments of FiO2 per 12 hour was 0 and 63, respectively.
CONCLUSION: Automated control of FiO2 significantly improved compliance of oxygen saturation targeting and significantly reduced exposure to hypoxaemia as well as hyperoxaemia. ©2015 Foundation Acta Paediatrica. Published by John Wiley & Sons Ltd.

Entities:  

Keywords:  Closed loop; Hyperoxaemia; Hypoxaemia; Oxygen saturation; Premature infants

Mesh:

Substances:

Year:  2015        PMID: 26194933     DOI: 10.1111/apa.13137

Source DB:  PubMed          Journal:  Acta Paediatr        ISSN: 0803-5253            Impact factor:   2.299


  10 in total

1.  Environmental or Nasal Cannula Supplemental Oxygen for Preterm Infants: A Randomized Cross-Over Trial.

Authors:  Colm P Travers; Waldemar A Carlo; Arie Nakhmani; Shweta Bhatia; Samuel J Gentle; VenkataNagaSai Apurupa Amperayani; Premananda Indic; Inmaculada Aban; Namasivayam Ambalavanan
Journal:  J Pediatr       Date:  2018-04-25       Impact factor: 4.406

2.  Clinical Evaluation of an Automatic Oxygen Control System for Premature Infants Receiving High-Flow Nasal Cannula for Respiratory Support: A Pilot Study.

Authors:  Xuefeng Hou; Akram Faqeeh; Ramak Amjad; John Pardalos; Roger Fales
Journal:  J Med Device       Date:  2022-05-10       Impact factor: 0.743

Review 3.  Automated Oxygen Delivery in Neonatal Intensive Care.

Authors:  Vrinda Nair; Prakash Loganathan; Mithilesh Kumar Lal; Thomas Bachman
Journal:  Front Pediatr       Date:  2022-06-22       Impact factor: 3.569

4.  Frequency and duration of extreme hypoxemic and hyperoxemic episodes during manual and automatic oxygen control in preterm infants: a retrospective cohort analysis from randomized studies.

Authors:  Thomas E Bachman; Wes Onland; Anton H van Kaam; Karel Roubik; Helmut D Hummler; Mithilesh Lal; Gianluca Lista; Carlos A Fajardo
Journal:  BMC Pediatr       Date:  2022-06-17       Impact factor: 2.567

Review 5.  Oxygen Saturation Targets in Preterm Infants and Outcomes at 18-24 Months: A Systematic Review.

Authors:  Veena Manja; Ola D Saugstad; Satyan Lakshminrusimha
Journal:  Pediatrics       Date:  2016-12-05       Impact factor: 7.124

6.  Eligibility Criteria and Representativeness of Randomized Clinical Trials That Include Infants Born Extremely Premature: A Systematic Review.

Authors:  Leeann R Pavlek; Brian K Rivera; Charles V Smith; Joanie Randle; Cory Hanlon; Kristi Small; Edward F Bell; Matthew A Rysavy; Sara Conroy; Carl H Backes
Journal:  J Pediatr       Date:  2021-04-21       Impact factor: 6.314

7.  Automated oxygen control with O2matic® during admission with exacerbation of COPD.

Authors:  Ejvind Frausing Hansen; Jens Dahlgaard Hove; Charlotte Sandau Bech; Jens-Ulrik Stæhr Jensen; Thomas Kallemose; Jørgen Vestbo
Journal:  Int J Chron Obstruct Pulmon Dis       Date:  2018-12-14

8.  Automated closed-loop FiO2 titration increases the percentage of time spent in optimal zones of oxygen saturation in pediatric patients-A randomized crossover clinical trial.

Authors:  Ekin Soydan; Gokhan Ceylan; Sevgi Topal; Pinar Hepduman; Gulhan Atakul; Mustafa Colak; Ozlem Sandal; Ferhat Sari; Utku Karaarslan; Dominik Novotni; Marcus J Schultz; Hasan Agin
Journal:  Front Med (Lausanne)       Date:  2022-08-25

9.  The effect of automated oxygen control on clinical outcomes in preterm infants: a pre- and post-implementation cohort study.

Authors:  H H Salverda; N J Oldenburger; M Rijken; S C Pauws; P A Dargaville; A B Te Pas
Journal:  Eur J Pediatr       Date:  2021-02-23       Impact factor: 3.183

10.  Comparison of two devices for automated oxygen control in preterm infants: a randomised crossover trial.

Authors:  Hylke H Salverda; Sophie J E Cramer; Ruben S G M Witlox; Timothy J Gale; Peter A Dargaville; Steffen C Pauws; Arjan B Te Pas
Journal:  Arch Dis Child Fetal Neonatal Ed       Date:  2021-06-10       Impact factor: 5.747

  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.