| Literature DB >> 26193410 |
Haijin Fu, Pengcheng Hu, Jiubin Tan, Zhigang Fan.
Abstract
A simple method was proposed by using a tunable attenuator fitted in the reference or measurement arm of a heterodyne laser interferometer to adjust the values of mixing laser beams while the spectrum of the measurement signal is monitored using a signal analyzer. The effectiveness of the proposed method in reducing the first-order optical nonlinearity was verified through experiments. Results indicated that the peak value of the first-order optical nonlinearity could be reduced from 5.15 to 0.24 nm. It was therefore concluded that the proposed method was applicable to ultraprecision laser interferometry.Year: 2015 PMID: 26193410 DOI: 10.1364/AO.54.006321
Source DB: PubMed Journal: Appl Opt ISSN: 1559-128X Impact factor: 1.980