Tiffany Guo1, Ritish Patnaik, Kevin Kuhlmann, Alex J Rai, Samuel K Sia.
Abstract
It is traditionally difficult to incorporate two classes of diagnostic tests into a single platform. In this work, we demonstrate a microfluidic-based smartphone dongle that simultaneously measures concentration of hemoglobin and detects HIV antibodies. Specifically, we demonstrate how a previously published immunoassay device, which measured optical density of silver precipitation on gold colloids, can be expanded to quantitatively measure hemoglobin concentration via a colorimetric assay. By lysing whole blood components with CHAPS detergent, we achieved highly reproducible measurement of hemoglobin concentration with the device. We tested this dual test on 38 patient samples from Columbia University Medical Center. Compared with the Hemocue Hb 201+ analyzer, hemoglobin concentrations from our device were accurate within 1.2 g dL(-1), while the HIV immunoassay (in the presence of CHAPS detergent) showed 95% sensitivity and 95% specificity, comparable to our previous studies. This work demonstrates the feasibility of integrating two classes of diagnostic tests (a colorimetric-based quantitative measurement and an immunoassay based on silver precipitation on gold colloids) into a low-cost, fast, and low-power dongle that works with smartphones, and creates a novel dual panel with clinical utility for antenatal-care settings.Entities:
Mesh:
Substances:
Year: 2015 PMID: 26190320 DOI: 10.1039/c5lc00609k
Source DB: PubMed Journal: Lab Chip ISSN: 1473-0189 Impact factor: 6.799