Pascal Meylan1, Luc Paris2, Oliver Liesenfeld3.
Abstract
Detection of IgG and IgM antibodies is commonly performed for the diagnosis of infection with Toxoplasma gondii. We determined the accuracy of the Elecsys Toxo IgG and IgM test at four European laboratories compared to local reference methods. Coefficients of variation for reproducibility ranged from 1.0 to 6.5% for IgG and from 0.8 to 3.2% for IgM. Seroconversion panels revealed high overall concordance with the reference tests. The Elecsys test detected IgG antibodies earlier than the Cobas Core IgG test in 19 of 47 panels; persisting IgM antibodies were observed in the VIDAS but not the Elecsys test in five of 47 panels. In 31.4% of latent stage sera with persistent IgM antibodies (positive LIASON IgM), the Elecsys IgM test gave negative results indicating increased "clinical" specificity. Sensitivity and specificity of the Elecsys IgG assay ranged from 99.45 to 100% and 87.50-99.80%, respectively, and 91.11-95.74 and 98.45-99.79% for the Elecsys IgM assay, respectively. In conclusion, excellent reproducibility and accuracy make the Elecsys Toxo G and M tests highly suitable for the detection of anti-T. gondii IgG and IgM antibodies. The lower detection rates for persistent IgM in the Elecsys IgM test increase "clinical" specificity and decrease the need for follow-up testing.Entities:
Keywords: IgG; IgM; Toxoplasma gondii; antibodies; latent infection; seroconversion
Year: 2015 PMID: 26185683 PMCID: PMC4500066 DOI: 10.1556/1886.2015.00016
Source DB: PubMed Journal: Eur J Microbiol Immunol (Bp) ISSN: 2062-509X
Discordant results in seroconversion panels (n = 25 in Marseille, n = 22 in Paris) between Elecsys and comparator IgG and IgM tests
| Result constellation Elecsys/comparator | IgG | IgM | |||
|---|---|---|---|---|---|
| Marseille (Elecsys vs. Cobas Core) | Paris (Elecsys vs. Platelia) | Marseille (Elecsys vs. VIDAS) | Paris (Elecsys vs. Platelia) | ||
| Seroconversion (negative to positive) | –/+ | 0 | 0 | 0 | 2 |
| ±/– | 8 | 8 | 0 | 0 | |
| +/– | 1 | 2 | 0 | 0 | |
| Disappearance of antibodies in f/up sera (positive to negative) | –/+ | 0 | 0 | 5 | 0 |
| ±/– | 0 | 0 | 0 | 0 | |
| +/– | 0 | 0 | 0 | 0 | |
*VIDAS IgM also gave negative results in the same sera
**After 4–36 weeks
Detection of anti-T. gondii IgM antibodies in patients with latent Toxoplasma infection (>3 month post infection as indicated by the presence of high IgG avidity antibodies)
| Serum no. | Liaison Toxo IgM | Elecsys Toxo IgM | Cobas Core IgM |
|---|---|---|---|
| 1 | 11.5 (+) | 0.45 | 47.70 |
| 2 | 37.0 (+) | 1.89 (+) | 251.50 (+) |
| 3 | 12.0 (+) | 1.59 (+) | 171.80 (+) |
| 4 | 12.0 (+) | 0.89 | 156.50 (+) |
| 5 | 80.0 (+) | 1.50 (+) | 274.0 (+) |
| 6 | 11.0 (+) | 0.43 | 30.80 |
| 7 | 14.0 (+) | 0.27 | 28.50 |
| 8 | 15.0 (+) | 1.48 (+) | 155.70 (+) |
| 9 | 11.0 (+) | 3.15 (+) | 291.30 (+) |
| 10 | 19.0 (+) | 2.59 (+) | 147.70 (+) |
| 11 | 18.0 (+) | 1.47 (+) | 192.50 (+) |
| 12 | 14.0 (+) | 0.75 | 94.30 |
| 13 | 14.0 (+) | 1.68 (+) | 90.80 |
| 14 | 39.0 (+) | 1.70 (+) | 399.40 (+) |
| 15 | 28.0 (+) | 1.88 (+) | 185.10 (+) |
| 16 | 75.0 (+) | 1.58 (+) | 407.20 (+) |
| 17 | 30.0 (+) | 0.22 | 10.30 |
| 18 | 19.0 (+) | 7.14 (+) | 564.70 (+) |
| 19 | 15.0 (+) | 3.75 (+) | 409.8 (+) |
| 20 | 14.0 (+) | 0.39 | 36.10 |
| 21 | 13.0 (+) | 0.24 | 13.70 |
| 22 | 16.0 (+) | 10.03 (+) | 564.70 (+) |
| 23 | 45.0 (+) | 3.45 (+) | 535.50 (+) |
| 24 | 15.0 (+) | 1.40 (+) | 228.80 (+) |
| 25 | 26.0 (+) | 2.95 (+) | 414.10 (+) |
| 26 | 11.0 (+) | 0.20 | 17.40 |
| 27 | 10.0 (+) | 1.89 (+) | 191.0 (+) |
| 28 | 119.0 (+) | 2.45 (+) | 397.7 (+) |
| 29 | 12.0 (+) | 1.82 (+) | 132.80 (+) |
| 30 | 22.0 (+) | 4.03 (+) | 564.70 (+) |
| 31 | 10.0 (+) | 1.22 (+) | 81.90 |
| 32 | 44.0 (+) | 0.26 | 23.40 |
| 33 | >150 (+) | 5.78 (+) | 564.70 (+) |
| 34 | 14.0 (+) | 3.94 (+) | 519.80 (+) |
| 35 | 42.0 (+) | 0.30 | 31.90 |
*(+), positive result
Relative sensitivity and relative specificity of the Elecsys Toxo G test compared to commercially available reference IgG assays before and after resolution of discordant results
| Site | Comparator assay | IgG before resolution | IgG after resolution | ||
|---|---|---|---|---|---|
| Sensitivity (CI) | Specificity (CI) | Sensitivity (CI) | Specificity (CI) | ||
| Berlin | VIDAS | 99.44 | 85.37 | 99.45 | 87.5 |
| Lausanne | VIDAS | 99.55 | 96.6 | 99.55 | 99.49 |
| Marseille | Cobas Core | 99.29 | 96.55 | 100 | 99. 57 |
| Paris | Platelia | 99.56 | 94.49 | 100 | 100 |
n.d., CI not determined. For specific results of discrepant testing, see the Results section of the article
Relative sensitivity and relative specificity of the Elecsys Toxo M test compared to commercially available IgM assays before and after resolution of discordant results using reference tests
| Site | Comparator assay | IgG before resolution | IgG after resolution | ||
|---|---|---|---|---|---|
| Sensitivity (CI) | Specificity (CI) | Sensitivity (CI) | Specificity (CI) | ||
| Berlin | VIDAS | 74.44 | 98.95 | 91.11 | 99.21 |
| Lausanne | VIDAS | 88.30 | 98.45 | 95.74 | 98.45 |
| Marseille | VIDAS | 87.20 | 98.82 | 93.13 | 99.11 |
| Paris | Platelia | 88.69 | 99.69 | 95.51 | 99.69 |