Early Performance and Safety of the Micra Transcatheter Pacemaker in Pigs.

Matthew Bonner1, Michael Eggen1, Tarek Haddad2, Todd Sheldon1, Eric Williams3.   

Abstract

BACKGROUND: The Micra® Transcatheter Pacing System (TPS; Medtronic Inc., Minneapolis, MN, USA) is a miniaturized single-chamber pacemaker system that is delivered via catheter through the femoral vein. In this study, the electrical performance was compared between the TPS and a traditional leaded pacemaker. In addition, the safety profile of the two systems was compared by thorough monitoring for a number of adverse events.
METHODS: The TPS was implanted in the right ventricular apex of 10 Yucatan mini pigs and a Medtronic single-lead pacemaker (SLP) was implanted in the right ventricular apex of another 10 pigs and connected to a traditional pacemaker. The electrical performance of all devices was monitored for 12 weeks. The safety profile of each system was characterized using x-ray, computed tomography, ultrasound, blood work, and necropsy to monitor for a variety of adverse events.
RESULTS: At implant the mean pacing thresholds were 0.58 ± 0.17 V @0.24 ms and 0.75 ± 0.29 V @0.21 ms for the TPS and the SLP respectively. After 12 weeks, mean thresholds were 0.94 ± 0.46 V and 1.85 ± 0.75 V (P < 0.0001). There were two pulmonary emboli that were small and past the tertiary branch, and one occurred in each arm. There were also two infections with one in each arm. There were no dislodgements (macro or micro), tissue injury, tamponade, or valve injury.
CONCLUSIONS: Overall, despite the 10-fold size reduction of the Micra TPS, it appears to perform similarly and have a similar safety profile to a traditional pacemaker system.
© 2015 Medtronic PLC. Pacing and Clinical Electrophysiology published by Wiley Periodicals, Inc.

Entities:  

Keywords:  miniaturization; new technology; pacemaker; transcatheter pacing system

Mesh:

Year:  2015        PMID: 26183288     DOI: 10.1111/pace.12694

Source DB:  PubMed          Journal:  Pacing Clin Electrophysiol        ISSN: 0147-8389            Impact factor:   1.976


  3 in total

1.  Cardiac resynchronization therapy with wireless left ventricular endocardial pacing: is this the direction to go?

Authors:  Tomas Holubec; Julius Beckers; Zdenka Holubcova; Veronika Walter; Thomas Walther
Journal:  Cardiovasc Diagn Ther       Date:  2018-08

2.  Right Ventricular Anatomy Can Accommodate Multiple Micra Transcatheter Pacemakers.

Authors:  Pamela Omdahl; Michael D Eggen; Matthew D Bonner; Paul A Iaizzo; Kent Wika
Journal:  Pacing Clin Electrophysiol       Date:  2016-02-01       Impact factor: 1.976

3.  Visualization and appearance of artifacts of leadless pacemaker systems in cardiac MRI : An experimental ex vivo study.

Authors:  Christoph Edlinger; Marcel Granitz; Vera Paar; Christian Jung; Alexander Pfeil; Sarah Eder; Bernhard Wernly; Jürgen Kammler; Klaus Hergan; Uta C Hoppe; Clemens Steinwender; Michael Lichtenauer; Alexander Kypta
Journal:  Wien Klin Wochenschr       Date:  2018-05-23       Impact factor: 1.704

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.