Reliability and responsiveness of algometry for measuring pressure pain threshold in patients with knee osteoarthritis.

Ebru Kaya Mutlu1, Arzu Razak Ozdincler1.   

Abstract

[Purpose] This study aimed to establish the intrarater reliability and responsiveness of a clinically available algometer in patients with knee osteoarthritis as well as to determine the minimum-detectable-change and standard error of measurement of testing to facilitate clinical interpretation of temporal changes. [Subjects] Seventy-three patients with knee osteoarthritis were included. [Methods] Pressure pain threshold measured by algometry was evaluated 3 times at 2-min intervals over 2 clinically relevant sites-mediolateral to the medial femoral tubercle (distal) and lateral to the medial malleolus (local)-on the same day. Intrarater reliability was estimated by intraclass correlation coefficients. The minimum-detectable-change and standard error of measurement were calculated. As a measure of responsiveness, the effect size was calculated for the results at baseline and after treatment.
[Results] The intrarater reliability was almost perfect (intraclass correlation coefficient = 0.93-0.97). The standard error of measurement and minimum-detectable-change were 0.70-0.66 and 1.62-1.53, respectively. The pressure pain threshold over the distal site was inadequately responsive in knee osteoarthritis, but the local site was responsive. The effect size was 0.70.
[Conclusion] Algometry is reliable and responsive to assess measures of pressure pain threshold for evaluating pain patients with knee osteoarthritis.

Entities:  

Keywords:  Pressure pain threshold; Reliability; Responsiveness

Year:  2015        PMID: 26180358      PMCID: PMC4500021          DOI: 10.1589/jpts.27.1961

Source DB:  PubMed          Journal:  J Phys Ther Sci        ISSN: 0915-5287


INTRODUCTION

Pain is inherently subjective, and pain measurement in patients with knee osteoarthritis (OA) relies primarily on self-reports1). The most common approaches to self-reported pain measurement are the use of a visual analog scale (VAS), numeric pain rating scales, and the Western Ontario and McMaster Asteoarthritis Index pain scale2,3,4). Although self-reported pain intensity is important, it is a composite of the physiological and psychological features of the patient and their health problem that is further mediated by social aspects, which can make it difficult to interpret responses5, 6). Thus, objective pain measures are invaluable as they reflect different perspectives of the health condition. Therefore, measuring knee pain is an important component of clinical practice; its importance is evident in the frequency with which it drives healthcare utilization as well as its impact on quality of life7). An important step toward integrating pressure pain threshold (PPT) testing into routine clinical practice is the establishment of the reliability of viable instruments8, 9). Such instruments must be commercially available, meet measurement standards under knee OA, and function under ideal conditions; furthermore, they must not be cost-prohibitive when used in a clinical setting. However, the clinimetrics of PPT testing with devices currently available in clinical practice are poorly understood for patients with knee OA. Pain threshold is evaluated by methods including cuff algometry, pressure algometry, and algometry with electric stimulation8, 9). Results in the literature suggest electronic and pressure algometers have comparable reliability10, 11). Unfortunately, the costs of electronic pressure algometers limit their use in routine clinical practice. However, pressure algometers are inexpensive, more convenient, and more widely available12). Moreover, pressure algometry methods can be used for clinical research to measure the efficacy of therapeutic interventions for the treatment of pain as well as general psychophysiological research8). The validity and reliability of self-rating scales such as the VAS for use in adults in general have been demonstrated13). However, research on the reliability of experimental pain measurement tests is limited. Studies of PPT have preliminarily concluded that it is a valid and reliable pain measurement tool in patients with knee OA14). However, the responsiveness, standard error of measurement (SEM), and minimum detectable change (MDC) have not been evaluated. Moreover, numerous studies in adults report differing intra- and interrater reliability of PPT among pathologies15, 16). In addition, the intrarater reliability of PPT measurements in patients with knee OA has not been sufficiently demonstrated. Therefore, this study aimed to establish the intrarater reliability and responsiveness of a clinically available algometer for patients with knee OA as well as to determine the MDC and SEM of testing to facilitate clinical interpretation of temporal changes.

SUBJECTS AND METHODS

Seventy-three patients (141 knees; mean age, 54.5 years; range, 40–69 years) with knee OA who received physical therapy for knee pain at Istanbul University, Faculty of Health Sciences, Division of Physiotherapy and Rehabilitation were included. Before the study, the patients completed an informed consent form that was approved by the Ethical Committee at Bakirkoy Dr. Sadi Konuk Education and Research Hospital (IRB study protocol: 201-146). The inclusion criterion was diagnosis of knee OA according to the American College of Rheumatology criteria and grade 2 or 3 knee OA according to the Kellgren and Lawrence (1957) scale17, 18). The exclusion criteria were rheumatoid arthritis; history of knee or hip joint replacement surgery of the affected joint; history of any other surgical procedure on the lower limbs in the previous 6 months; a planned surgical procedure on the lower limbs in the next 6 months; the initiation of opioid analgesia, or corticosteroid or analgesic injection interventions for knee pain within the previous 3 months; and any physical therapy intervention on the lower limbs in the previous 6 months. We analyzed variables including age, gender, symptom duration, and pain intensity. Mechanical pain was evaluated using an algometer. Pain intensity was assessed using a VAS in which the patient is asked to indicate his/her perceived pain during rest, activity, and nighttime; pain is scored from 0 to 10 on a numeric pain rating scale, with 0 indicating no pain and 10 indicating the worst pain possible19). The patients (n = 83) attended 12 treatment sessions 3 times per week in the physical therapy clinic and underwent assessments before treatment and after 4 weeks of treatment. However, 10 patients did not complete the study (drop-out rate: 13.6%): 5 females and 1 male dropped out because of changes in their work situation, and the other 2 females and 2 males dropped out because of other health problems. The Baseline 1200-304 (Push-Pull Force Gauge®, Fabrication Enterprises, Inc.) is a handheld pressure algometer that responds linearly to force application between 0 and 10 kg (22 lb × 0.25 lb and 10 kg × 100 g); it has a 1-cm2 round rubber tip, and values are displayed as the maximum force applied before the individual verbally states that the pain threshold has been reached. The following standardized method was used when evaluating the patientsPPT. First, 1 PPT measurement trial was performed with the algometer on the hand to ensure the measurement procedure was understood; each patient was coached how to differentiate his or her report of tactile and painful stimuli. The PPT was recorded as the amount of pressure required to elicit a sensation of pain distinct from pressure or discomfort20). We asked the patients to say “stop” as soon as a discernible sensation of pain was felt; at this point, the algometer pressure was immediately released, and the plunger was retracted by the rater. The algometer probe was lowered at a constant rate of approximately 1 lb/s until the PPT was reached, as indicated by the patient’s verbal notification. In the present study, the rater was a female physical therapist. All PPT measurements were applied by the same examiner, and the rate of algometric pressure application was constant to ensure good reliability. The algometric measurements were performed while the patient was in the side-lying position according to the study by Moss et al21). The localization of the knee and ankle algometric pressure is easier with the patient lying on his/her side than in the supine position, especially in obese patients. In addition, holding the algometer vertically as opposed to horizontally is advantageous when applying the pressure21). The PPT was subsequently assessed in all patients at 2 sites by one rater: the medial aspect of the patient’s affected knee (approximately 1–2 cm mediolateral to the medial femoral tubercle) and the medial heel (approximately 1–2 cm lateral to the medial malleolus). These body sites were chosen, because they are painful and pain-free areas, respectively, in knee OA patients. With the patient lying on his/her side, a 1-cm2 algometer probe was used to apply pressure to the skin at 90°21). The PPT was measured 3 times at each site on each side of the body, first starting at the ankle and then moving to the knee. Measurements were made approximately every 2 min. For consistency, the right side was tested first. Temperature, noise, and ambient lighting were kept as constant as possible throughout the testing sessions. Walter et al.22) developed a robust mathematical approach for estimating the required number of patients for reliability studies. The hypothesis is that the test-retest reliability is lower than the interrater reliability for an estimated intraclass correlation coefficient (ICC) of 0.8 with a null hypothesis level set at 0.4. Using these parameters, a sample of 27 patients provides 80% statistical power with 95% confidence (p < 0.05). Concordantly, previous studies assessing the reliability of PPT in other pathologies have included between 13 and 70 patients11, 14, 16). Therefore, a sample size of 70 knee OA patients was assumed adequate to assess the reliability of PPT in the present study. All analyses were conducted using SPSS version 21. The level of significance was set at p < 0.05. Prior to determining statistical associations, the Kolmogorov-Smirnov test was performed to assess the distribution of data; all data were normally distributed. The data are expressed as means ± standard deviations. The sample size adhered to the parameters and conditions governing parametric tests. Intragroup comparisons before and after treatment were performed by the paired t-test. Many previous studies of the reliability of equipment-based measurements use Pearson correlation coefficients for statistical analysis20). However, ICCs are arguably a more appropriate method for analyzing intrarater reliability23). The calculation of 95% confidence intervals to identify the precision of the estimate is also recommended together with the SEM, which is a measure of the (im)precision of the measurements themselves24, 25). Therefore, in the present study, ICCs were used to estimate intrarater reliability. Estimates of reliability were calculated using relative (i.e., ICC for absolute agreement) and absolute (i.e., SEM and MDC) estimates23). Reliability coefficients were interpreted with the subjective categories of Landis and Koch26) as follows: <0.40, 0.41–0.60, 0.61–0.80, and 0.81–1.00 correspond to unacceptable, moderate, substantial, and almost perfect agreement, respectively. A large ICC indicates that the test is objective and standardized23). Error in the instrument or method would induce bias and low repeatability. The SEM, as an indicator of the expected measurement error in an individual’s score using the same units as the algometer, was calculated as follows: SDpooled × √1 − ICC24). The SEM represents the extent to which a patient’s performance measurements vary if the test is repeated without any underlying change in the patient; in other words, it represents measurement error24). The MDC was calculated at the 90% level, which is appropriate for assessing change during routine clinical use25). The threshold amount of change in scores required for the rater to be 90% confident that true change exceeding the measurement error had occurred was determined by the following formula: MDC90 = SEM × √2 × 1.6427). The MDC indicates the amount of change required to exceed measurement variability25); in other words, it represents the smallest change of an outcome measure that would be considered “real”. As a measure of responsiveness, the effect size was calculated for the results of the evaluations at baseline and after 4 weeks. To evaluate the changes due to physiotherapy, we used data from the patients who completed both evaluations. The effect size was calculated as the mean score difference divided by the standard deviation from the initial measurement as described by Kazis et al28). An effect size >0.80 is considered high.

RESULTS

The study patients (n = 73) were predominantly female (84.9%) and had a mean symptom duration of 8 months (Table 1). Pain (i.e., VAS-rest, VAS-activity, and VAS-night) significantly reduced after treatment (p < 0.05) (Table 1). The intrarater reliability between test sessions 1 and 2 before and after treatment on the medial knee were excellent (ICC1–2 = 0.95 and ICC1–2 = 0.96, respectively). The intrarater reliability was also excellent at both sites between testing times 1 and 3, and 2 and 3 (all p < 0.001) (Table 2). The SEM and MDC were 0.66 and 1.53 for the medial knee, and 0.70 and 1.62 for the medial malleolus, respectively (Table 3).
Table 1.

Patient characteristics

Before treatment (n=83)After treatment (n=73)
Gender (F/M)62/1155/8-
Age (years)55.5±6.756.24 ±6.7-
BMI31.5±4.931.39 ±5.1-
Duration of symptoms (months)8.1±10.67.6 ±10.7-
VAS-rest3.7 ±2.81.3 ±2.1***
VAS-activity7.8 ±2.24.7 ±2.9***
VAS-night5.2 ±3.71.8 ±3.03***

Data are mean ± SD. BMI: body mass index; VAS: visual analogue scale. ***p < 0.001, paired t-test

Table 2.

Measures and intrarater reliability (ICC) of PPT testing using an algometer at each site in knee osteoarthritis

PPT-test (LBS)Measure- Mean±SDICC Comparisons


Test 1Test 2Test 3ICC1,2 (95%CI)ICC2,3 (95%CI)ICC1,3 (95%CI)
Before Treatment
Medial malleol7.65±2.797.38±2.627.48±2.700.93 (0.90–0.95)0.97 (0.95–0.98)0.94 (0.90–0.96)
Medial knee5.54±2.975.41±3.105.46±3.010.95 (0.94–0.96)0.97 (0.94–0.98)0.96 (0.94–0.97)
After Treatment
Medial malleol8.19±2.698.24±2.78.18±2.660.97 (0.96–0.98)0.97 (0.96–0.98)0.98 (0.97–0.98)
Medial knee7.51±2.927.55±3.017.68±3.060.96 (0.94–0.97)0.97 (0.95–0.98)0.96 (0.95–0.97)
Table 3.

Standard error of measurement (SEM), minimum detectable change (MDC), and effect size (ES) of PPT testing (Units: lb)

Before treatmentMeasure- Mean (SD)After treatmentMeasure- Mean (SD)SEMMDCES
Medial malleol7.50±2.668.20±2.670.701.260.26
Medial knee5.47±2.997.58±2.970.661.190.70
Data are mean ± SD. BMI: body mass index; VAS: visual analogue scale. ***p < 0.001, paired t-test The mean PPTs at the medial malleolus and knee sites before treatment were 7.50 ± 2.66 and 5.47 ± 2.99, respectively; those after treatment were 8.20 ± 2.67 and 7.58 ± 2.97, respectively (Table 3). The effect size data are shown in Table 3.

DISCUSSION

The results of the present study suggest that PPT testing is reliable and acceptable irrespective of the presence of knee pain in the patient undergoing testing. The intrasession repeatability of a single PPT measurement taken by the same rater was excellent for both the knees and ankles in patients with knee OA. The absolute error, reflected by the SEM and MDC, indicates that the clinical measurement properties of the PPT evaluated in this study were appropriate. In addition, our findings show that the PPT is responsive to a physical therapeutic intervention. Our results are comparable to those of other investigations. Persson et al. report intrarater ICCs ranging from 0.70 to 0.90 for testing the upper trapezius muscles of 27 healthy female patients29). Furthermore, Paungmali et al.15) performed PPT testing over the local lumbar area; their intrarater ICC of 0.99 is comparable to our ICC for the knee. Wessel14) reports that the reliability of pain threshold varies from 0.61 to 0.91 between repeated measurements in knee OA. Moreover, Wylde et al.30) report no significant differences in the PPTs at any body site in OA patients between baseline and 1 week, with high or very high ICCs (0.83–0.91). In addition, Moss et al.21) report that the ICCs (95% confidence interval) for pilot data (n = 5) indicated reasonable intrasubject reliability for both knee PPT (ICC = 0.94 [0.55–0.99] and heel PPT (ICC = 0.94 [0.59–0.99]). The present study found high reliability (ICC = 0.93–0.97) comparable to that in the previous studies. Therefore, our findings suggest that only one PPT measurement needs to be performed, because reliability of this protocol produced high ICCs for the same examiner. Paungmali et al.15) report an SEM of 1.19 for testing over the local lumbar area of patients with low back pain; meanwhile, the SEM in the present study was 0.66. This suggests that the local lumbar area muscles may be a more labile site with respect to pain threshold than the knee. In the present study, the absolute error, reflected by the SEM and MDC, suggests that the clinical measurement properties were appropriate. Clinicians will be able to use these values in their communications with funders or other healthcare professionals to indicate their level of confidence in the extent and meaning of score changes in their patients. However, clinicians should note that the MDC at the knee site of 1.53 indicates that the tool may not be sufficiently accurate to detect a decrease in the PPT if the baseline value is smaller than the MDC, as was the case in patients with knee OA. Physical therapy is effective for pain relief and improvement of quality of life in patients with knee OA31). However, the effect size is large, suggesting that the PPT is suitable for evaluating the effects of a physical therapeutic intervention28). Walton et al.32) show that the PPT over the upper trapezius muscle is responsive and useful as part of a protocol to evaluate clinical changes. The results of the present study indicate that patients with knee OA who received physical therapy experienced clinically meaningful improvements in pain as evidenced by the effect sizes for pain. The effect size was moderate at 0.70, indicating the PPT is sufficiently sensitive for detecting changes over time. Algometry is easy to perform in clinical settings9), as equipment, training, time, and physical space requirements are minimal. Algometry has been demonstrated to be adequately reliable for research when high-precision instruments are used in healthy subjects or patients33,34,35). However, Arendt-Nielsen et al.36) and Neogi et al.37) report that knee OA duration and radiographic findings are not associated with the PPT. On the other hand, Imamura et al.38) and Lee et al.39) demonstrate that OA patients have lower PPTs than controls across multiple body sites. Also, Wyle et al.30) support the inclusion of pressure algometry in studies assessing pain perception abnormalities in OA. Thus, it can be concluded that the PPT can be used in research. In adults, PPTs decrease with age40), and females exhibit lower thresholds than males41). Age may also influence sensitivity to pressure pain independent of the disease progression, as evidenced by an increase in the PPT with age42). Other factors such as the patient’s levels of anxiety and anticipation may affect response to pressure pain induced by a pressure algometer43). In addition, studies of adults report experimenter gender and professional status may influence thresholds to pain stimuli44, 45). In a study by Kallai et al.44), 160 healthy participants were asked to immerse their hand in cold water (−1 °C) as long as possible. The participants tolerated pain longer when examined by a faculty member than when examined by a student. In addition, the participants expressed higher pain intensity when examined by a female rater. The present study was performed by professional rater, which is a strength of the study. The major limitation of this study is that we could not assess interrater reliability, which is a critical measure for evaluating the PPT. Although the present study demonstrated the reliability and responsiveness of the PPT, these values should be corroborated by studies with larger populations. The second major limitation is that we selected only one pain-sensitive site from the knee joint area. In addition, we did not ask the patients if they perceived a decrease in their PPT. Therefore, additional studies are required to determine the minimum clinically important differences in the PPT in patients with knee OA. The present study indicates that using algometry to measure the PPT is reliable and responsive, and can be used in patients with knee OA in clinical practice, especially in a single measurement session performed by a professional examiner.
  42 in total

1.  Radiological assessment of osteo-arthrosis.

Authors:  J H KELLGREN; J S LAWRENCE
Journal:  Ann Rheum Dis       Date:  1957-12       Impact factor: 19.103

2.  Reliability and validity of a pressure algometer.

Authors:  Ann M Kinser; William A Sands; Michael H Stone
Journal:  J Strength Cond Res       Date:  2009-01       Impact factor: 3.775

Review 3.  Toward a biopsychomotor conceptualization of pain: implications for research and intervention.

Authors:  Michael J L Sullivan
Journal:  Clin J Pain       Date:  2008-05       Impact factor: 3.442

4.  Test-retest reliability, repeatability, and sensitivity of an automated deformation-controlled indentation on pressure pain threshold measurement.

Authors:  Terry K Koo; Jing-yi Guo; Cameron M Brown
Journal:  J Manipulative Physiol Ther       Date:  2013-02       Impact factor: 1.437

5.  Age effects on pain thresholds, temporal summation and spatial summation of heat and pressure pain.

Authors:  Stefan Lautenbacher; Miriam Kunz; Peter Strate; Jesper Nielsen; Lars Arendt-Nielsen
Journal:  Pain       Date:  2005-06       Impact factor: 6.961

6.  Clinical pressure pain threshold testing in neck pain: comparing protocols, responsiveness, and association with psychological variables.

Authors:  David M Walton; Lenerdene Levesque; Martin Payne; Julie Schick
Journal:  Phys Ther       Date:  2014-02-20

7.  Measuring mechanical pain: the refinement and standardization of pressure pain threshold measurements.

Authors:  Michael Melia; Martin Schmidt; Britta Geissler; Jochem König; Ulrike Krahn; Hans Jürgen Ottersbach; Stephan Letzel; Axel Muttray
Journal:  Behav Res Methods       Date:  2015-03

8.  Interrater reliability of algometry in measuring pressure pain thresholds in healthy humans, using multiple raters.

Authors:  Linda S Chesterton; Julius Sim; Christine C Wright; Nadine E Foster
Journal:  Clin J Pain       Date:  2007 Nov-Dec       Impact factor: 3.442

9.  How much change is true change? The minimum detectable change of the Berg Balance Scale in elderly people.

Authors:  Declan Donoghue; Emma K Stokes
Journal:  J Rehabil Med       Date:  2009-04       Impact factor: 2.912

10.  The reliability and validity of pain threshold measurements in osteoarthritis of the knee.

Authors:  J Wessel
Journal:  Scand J Rheumatol       Date:  1995       Impact factor: 3.641

View more
  16 in total

1.  Pressure algometry is a suitable tool to assess anterior knee pain in osteoarthritic patients.

Authors:  Pedro Hinarejos; Nerea Goicoechea; Mauricio Gidi; Joan Leal-Blanquet; Raul Torres-Claramunt; Juan Sánchez-Soler; Joan Carles Monllau
Journal:  Eur J Orthop Surg Traumatol       Date:  2019-02-07

2.  A clinical perspective on a pain neuroscience education approach to manual therapy.

Authors:  Adriaan Louw; Jo Nijs; Emilio J Puentedura
Journal:  J Man Manip Ther       Date:  2017-05-22

Review 3.  Quantitative sensory testing: identifying pain characteristics in patients with osteoarthritis.

Authors:  K R Arant; J N Katz; T Neogi
Journal:  Osteoarthritis Cartilage       Date:  2021-09-28       Impact factor: 6.576

4.  Association of neuropathic pain with ultrasonographic measurements of femoral cartilage thickness and clinical parameters in patients with knee osteoarthritis.

Authors:  Nilgün Mesci; Erkan Mesci; Duygu Geler Külcü
Journal:  J Phys Ther Sci       Date:  2016-08-31

5.  Traits associated with central pain augmentation in the Knee Pain In the Community (KPIC) cohort.

Authors:  Kehinde Akin-Akinyosoye; Nadia Frowd; Laura Marshall; Joanne Stocks; Gwen S Fernandes; Ana Valdes; Daniel F McWilliams; Weiya Zhang; Michael Doherty; Eamonn Ferguson; David A Walsh
Journal:  Pain       Date:  2018-06       Impact factor: 6.961

6.  Effect of Dynamic Balance Exercises Based on Visual Feedback on Physical Function, Balance Ability, and Depression in Women after Bilateral Total Knee Arthroplasty: A Randomized Controlled Trial.

Authors:  Ju-Yeon Lee; Jung-Hee Kim; Byoung-Hee Lee
Journal:  Int J Environ Res Public Health       Date:  2020-05-05       Impact factor: 3.390

7.  Effectiveness between Dry Needling and Ischemic Compression in the Triceps Surae Latent Myofascial Trigger Points of Triathletes on Pressure Pain Threshold and Thermography: A Single Blinded Randomized Clinical Trial.

Authors:  María Benito-de-Pedro; Ricardo Becerro-de-Bengoa-Vallejo; Marta Elena Losa-Iglesias; David Rodríguez-Sanz; Daniel López-López; Julia Cosín-Matamoros; Eva María Martínez-Jiménez; César Calvo-Lobo
Journal:  J Clin Med       Date:  2019-10-05       Impact factor: 4.241

8.  Effects of Diacutaneous Fibrolysis on Passive Neuromuscular Response and Mechanosensitivity in Athletes with Hamstring Shortening: A Randomized Controlled Trial.

Authors:  Aida Cadellans-Arróniz; Carlos López-de-Celis; Albert Pérez-Bellmunt; Jacobo Rodríguez-Sanz; Luis Llurda-Almuzara; Vanessa González-Rueda; Pere Ramón Rodríguez-Rubio
Journal:  Int J Environ Res Public Health       Date:  2021-06-18       Impact factor: 3.390

9.  Is One Trial Sufficient to Obtain Excellent Pressure Pain Threshold Reliability in the Low Back of Asymptomatic Individuals? A Test-Retest Study.

Authors:  Romain Balaguier; Pascal Madeleine; Nicolas Vuillerme
Journal:  PLoS One       Date:  2016-08-11       Impact factor: 3.240

10.  Patients Using an Online Forum for Reporting Progress When Engaging With a Six-Week Exercise Program for Knee Conditioning: Feasibility Study.

Authors:  Philip Bright; Karen Hambly
Journal:  JMIR Rehabil Assist Technol       Date:  2018-04-26
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.