The impact of noninvasive, capillary, and venous hemoglobin screening on donor deferrals and the hemoglobin content of red blood cells concentrates: a prospective study.

Ariane Sümnig1, Gregor Hron2, Antje Westphal1, Astrid Petersmann3, Thomas Kohlmann4, Andreas Greinacher1, Thomas Thiele1.   

Abstract

BACKGROUND: Hemoglobin (Hb) is screened before whole blood donation to protect donors from anemia. Recently, noninvasive methods have become available for Hb screening in blood donors. We compared a noninvasive, a capillary, and a venous method for Hb screening of blood donors. STUDY DESIGN AND METHODS: Consecutive donors were prospectively screened using a noninvasive (Masimo Pronto-7), a capillary (HemoCue Hb 301), and a venipuncture-based method as gold standard (Siemens Advia 2120i) for Hb determination. All measurements were performed in parallel and in duplicate. A cutoff Hb value of 125 g/L (females) and 135 g/L (males) was used for donor acceptance.
RESULTS: A total of 553 donors were analyzed; in 38 donors (7%) the noninvasive Hb method was not applicable due to technical reasons. The noninvasive test underestimated (mean bias, -5.9 g/L; 95% limits of agreement, -25.74, 13.88) and the capillary test overestimated Hb values (mean bias, 4.3 g/L; 95% limits of agreement, -8.13, 16.71). Coefficients of variation of duplicate measurements were 1.05 (venous), 2.73 (noninvasive), and 3.23 (capillary). The noninvasive test revealed false low Hb values in 21.2% and the capillary test revealed false high Hb values in 9% of donors compared to the venous method. The negative predictive value of the noninvasive test was 94.3%.
CONCLUSION: The noninvasive Hb measurement is a reasonable first-line approach for predonation Hb screening of blood donors but a second method should be available to retest those not testable with the noninvasive device or with Hb values below the cutoffs.
© 2015 AABB.

Entities:  

Mesh:

Substances:

Year:  2015        PMID: 26177874     DOI: 10.1111/trf.13241

Source DB:  PubMed          Journal:  Transfusion        ISSN: 0041-1132            Impact factor:   3.157


  7 in total

Review 1.  Methods and analyzers for hemoglobin measurement in clinical laboratories and field settings.

Authors:  Ralph D Whitehead; Zuguo Mei; Carine Mapango; Maria Elena D Jefferds
Journal:  Ann N Y Acad Sci       Date:  2019-06-04       Impact factor: 5.691

Review 2.  Emerging point-of-care technologies for anemia detection.

Authors:  Ran An; Yuning Huang; Yuncheng Man; Russell W Valentine; Erdem Kucukal; Utku Goreke; Zoe Sekyonda; Connie Piccone; Amma Owusu-Ansah; Sanjay Ahuja; Jane A Little; Umut A Gurkan
Journal:  Lab Chip       Date:  2021-05-18       Impact factor: 6.799

Review 3.  Techniques used for the screening of hemoglobin levels in blood donors: current insights and future directions.

Authors:  Rajendra Chaudhary; Anju Dubey; Atul Sonker
Journal:  J Blood Med       Date:  2017-07-03

4.  Comparison of four methods to measure haemoglobin concentrations in whole blood donors (COMPARE): A diagnostic accuracy study.

Authors:  Steven Bell; Michael Sweeting; Anna Ramond; Ryan Chung; Stephen Kaptoge; Matthew Walker; Thomas Bolton; Jennifer Sambrook; Carmel Moore; Amy McMahon; Sarah Fahle; Donna Cullen; Susan Mehenny; Angela M Wood; Jane Armitage; Willem H Ouwehand; Gail Miflin; David J Roberts; John Danesh; Emanuele Di Angelantonio
Journal:  Transfus Med       Date:  2020-12-20       Impact factor: 2.019

5.  Accuracy of Capillary Hemoglobin Measurements for the Detection of Anemia among U.S. Low-Income Toddlers and Pregnant Women.

Authors:  Safia Boghani; Zuguo Mei; Geraldine S Perry; Gary M Brittenham; Mary E Cogswell
Journal:  Nutrients       Date:  2017-03-09       Impact factor: 5.717

6.  Evaluation of new non-invasive & conventional invasive methods of haemoglobin estimation in blood donors.

Authors:  Diptiranjan Rout; Suchet Sachdev; Neelam Marwaha
Journal:  Indian J Med Res       Date:  2019-06       Impact factor: 2.375

7.  Digital Hemoglobinometers as Point-of-Care Testing Devices for Hemoglobin Estimation: A Validation Study from India.

Authors:  Kapil Yadav; Shashi Kant; Gomathi Ramaswamy; Farhad Ahamed; Kashish Vohra
Journal:  Indian J Community Med       Date:  2020-10-28
  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.