BACKGROUND: Endovascular aortic sealing (EVAS) with a sac anchoring endoprosthesis excludes abdominal aortic aneurysms based on polymer filling of endobags. Primary objective was to assess the reliability of pre-procedural computed tomography (CT) scans based calculations of required endobag volume in relation to intraoperative volume of the endobags.
METHODS: Forty elective EVAS patients were included. Pre-procedural estimations of endobag volume were based on CT segmentations of aortic flow lumen volume, including both automated and manually-adjusted segmentations, performed by two experienced users. Additionally, changes in maximum AAA diameter, thrombus volume and total AAA volume were calculated from pre- and post-procedural CT scans.
RESULTS: Automatically determined volumes were comparable to manually-adjusted calculations (75.3 vs. 75.7 mL) and inter-observer agreement regarding pre-EVAS calculations of prefill volume appeared almost perfect with an intra-class correlation coefficient of 0.98 (95% CI: 0.96-0.99). The mean pressure of the endobags was 185 mmHg. Manually-adjusted pre-procedural volume calculations underestimated procedural volume of the endobags (-11.3±9.9 mL). Differences between pre-EVAS and procedural volume measurements were independent from endobag pressure (r=-0.06, P=0.72), prepocedural thrombus volume (r=-0.303, P=0.057) and changes in total AAA volume (r=0.02, P=0.91). A significant association was determined between differences in pre-EVAS and endobag volume versus changes in thrombus volume pre- and post-procedural (r=0.39, P=0.01).
CONCLUSIONS: In this validation study, pre-procedural volume measurements underestimate the actual fill volume of the endobags. It should be advised to perform a prefill of the endobags during the EVAS procedure.

Entities:  

Mesh:

Year:  2015        PMID: 26173392     DOI: 10.23736/S0021-9509.16.08820-0

Source DB:  PubMed          Journal:  J Cardiovasc Surg (Torino)        ISSN: 0021-9509            Impact factor:   1.888


  2 in total

1.  Two-Year Outcomes of the Nellix EndoVascular Aneurysm Sealing System for Treatment of Abdominal Aortic Aneurysms.

Authors:  Aleksandra C Zoethout; Johannes T Boersen; Jan M M Heyligers; Jean-Paul P M de Vries; Clark J A M Zeebregts; Michel M P J Reijnen
Journal:  J Endovasc Ther       Date:  2018-03-29       Impact factor: 3.487

2.  Secondary Fill Minimizes Gutter Size in Chimney EVAS Configurations In Vitro.

Authors:  Theodorus G van Schaik; Jorn P Meekel; Vincent Jongkind; Rutger J Lely; Maarten Truijers; Arjan W J Hoksbergen; Willem Wisselink; Jan D Blankensteijn; Kak Khee Yeung
Journal:  J Endovasc Ther       Date:  2018-12-21       Impact factor: 3.487

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.