| Literature DB >> 26172424 |
Aurélien Daynès1,2, Nevzat Temurok1, Jean-Philippe Gineys1, Gilles Cauet1, Philippe Nerin1, Jean Baudry2, Jérôme Bibette2.
Abstract
We present the principle of a fast magnetic field enhanced colloidal agglutination assay, which is based on the acceleration of the recognition rate between ligands and receptors induced by magnetic forces. By applying a homogeneous magnetic field of 20 mT for only 7 s, we detect CRP (C-reactive protein) in human serum at a concentration as low as 1 pM for a total cycle time of about 1 min in a prototype analyzer. Such a short measurement time does not impair the performances of the assay when compared to longer experiments. The concentration range dynamic is shown to cover 3 orders of magnitude. An analytical model of agglutination is also successfully fitting our data obtained with a short magnetic pulse.Entities:
Mesh:
Substances:
Year: 2015 PMID: 26172424 DOI: 10.1021/acs.analchem.5b00279
Source DB: PubMed Journal: Anal Chem ISSN: 0003-2700 Impact factor: 6.986