Extremely Low Frequency Electromagnetic Field from Convective Air Warming System on Temperature Selection and Distance.

Kwang Rae Cho1, Myoung-Hun Kim1, Myoung Jin Ko2, Jae Wook Jung2, Ki Hwa Lee2, Yei-Heum Park2, Yong Han Kim2, Ki Hoon Kim3, Jin Soo Kim3.   

Abstract

BACKGROUND: Hypothermia generates potentially severe complications in operating or recovery room. Forced air warmer is effective to maintain body temperature. Extremely low frequency electromagnetic field (ELF-EMF) is harmful to human body and mainly produced by electronic equipment including convective air warming system. We investigated ELF-EMF from convective air warming device on various temperature selection and distance for guideline to protect medical personnel and patients.
METHODS: The intensity of ELF-EMF was measured as two-second interval for five minutes on various distance (0.1, 0.2, 0.3, 0.5 and 1meter) and temperature selection (high, medium, low and ambient). All of electrical devices were off including lamp, computer and air conditioner. Groups were compared using one-way ANOVA. P<0.05 was considered significant.
RESULTS: Mean values of ELF-EMF on the distance of 30 cm were 18.63, 18.44, 18.23 and 17.92 milligauss (mG) respectively (high, medium, low and ambient temperature set). ELF-EMF of high temperature set was higher than data of medium, low and ambient set in all the distances.
CONCLUSION: ELF-EMF from convective air warming system is higher in condition of more close location and higher temperature. ELF-EMF within thirty centimeters exceeds 2mG recommended by Swedish TCO guideline.

Entities:  

Keywords:  Air warmer; ELF-EMF; Operating room

Year:  2014        PMID: 26171355      PMCID: PMC4499084     

Source DB:  PubMed          Journal:  Iran J Public Health        ISSN: 2251-6085            Impact factor:   1.429


Introduction

Electromagnetic field exists everywhere electricity flows. The concerns about the hazard of electromagnetic field were first reported by Wertheimer and Leeper (1). They issued high prevalence of leukemia in children near electrical power lines compared to control children. Extremely low frequency electromagnetic fields (ELF-EMF) are produced by electrical currents that have wavelength of 50 or 60 Hz. All the electronic devices are operated with input of electricity that produces electromagnetic waves. After report of Wertheimer and Leeper, the carcinogenic effect of ELF-EMF was brought up on various kinds of cancer like brain, breast, and leukemia (2–4). The harmful effect of extremely low frequency electromagnetic field (ELF-EMF) to human body is not entirely understood. However, ELF-EMF can be hazardous to health in animal and in vitro study. Health risk was also documented in human based epidemiology. Exposure to high intensity of ELF-EMF is proved as relatively detrimental. Long-term effect of low intensity is still under the controversy (5). Hypothermia is one of most important problem during or after surgery. Perioperative hypothermia results from thermo-regulatary impairment or cold operating environment. Inadvertent hypothermia is associated with high mortality rate and complication. Convective air warming system is an effective method to maintain body temperature. ELFEMF was produced by Convective air warming unit like other electronic devices. Medical workers have a potential risk of exposure to ELF-EMF especially in case of long-time use. Though it is short-time use, repetitive exposure could be also harmful to medical staff. Air warming device (WarmTouchTM, Covidien, USA) has various target temperature on demand (Fig. 1).
Fig. 1:

WarmTouchTM manufactured by Covidien

WarmTouchTM manufactured by Covidien We investigated ELF-EMF from convective air warming device on various temperature selection and distance for guideline to protect medical personnel and patients.

Methods

This test was designed as prospective experimental study and conducted in an empty operating room of university hospital. The operating room is rectangular shape and 9.25 X 5.32 square meters. All the electrical appliances including computer, astral lamp, monitor, surgical and anesthetic devices had been off during measurement. Atmosphere temperature was 21°C. ELF-EMF meter takes a measurement of triple axis (X, Y, Z) from 2.5 times per second sampling rate. The intensity of ELF-EMF was measured as two-second interval for five minutes by ELF-EMF meter (TM-192D, Tenmars, Taiwan). Each group consists of 150 data on four-temperature selection (high, medium, low and ambient) and five different distances (0.1, 0.2, 0.3, 0.5 and 1 meter). All the data were stored in the device that operated by battery. Sample size was calculated by MedCalc version 11.6 as following conditions: type I error-alpha 0.05, type II error-beta 0.20, difference 0.01. Minimal required sample size was computed to 63. Each group was compared by One-Way ANOVA using SAS 9.2 (SAS Institute, Cary, NC). Post-hoc test was conducted with Tukey procedure. P<0.05 was considered significant.

Results

The mean value of ELF-EMF was posted in Table 1 on four-temperature selection (high, medium, low and ambient) and five different distances (0.1, 0.2, 0.3, 0.5 and 1 meter). Average ELF-EMF of high temperature selection was 134.706, 41.518, 18.637, 5.343 and 0.780 mG respectively (0.1, 0.2, 0.3, 0.5 and 1 meter) (Fig. 2).
Table 1:

ELF-EMF on various temperature and distance

TemperatureHighMediumLowAmbientAverage
Distance
0.1 meter134.706±0.448132.061±0.626131.760±0.927130.329±0.369132.214
0.2 meter41.518±0.54440.995±0.14440.705±0.23439.912±0.12640.783
0.3 meter18.637±0.28818.447±0.05518.239±0.07717.928±0.06018.313
0.5 meter5.343±0.0465.249±0.0335.230±0.0455.170±0.0375.248
1 meter0.780±0.0170.776±0.0160.774±0.0140.762±0.0120.773
Average40.07039.38539.22338.820

Data was presented as mean ± standard deviation.

Fig. 2:

ELF-EMF of high temperature set

ELF-EMF of high temperature set ELF-EMF on various temperature and distance Data was presented as mean ± standard deviation. ELF-EMF from long distance represented lower value in all kinds of temperature. Mean value of 0.3 meter distance was 18.637, 18.447, 18.239 and 17.928 mG respectively (high, medium, low and ambient temperature) over Swedish TCO guideline 2mG (Fig. 3). Measured values from the distance of 1 meter were under Swedish TCO guideline.
Fig. 3:

ELF-EMF on the distance of 30cm

ELF-EMF on the distance of 30cm The intensity of ELF-EMF was higher in high temperature set than in medium, low and ambient. All the groups were significantly different (P<0.05).

Discussion

ELF-EMF was classified as group 2B, possibly carcinogenic by international agency for research on cancer (IARC) that ELF-EMF could be some hazardous to human body (6). ELF-EMF may be produced in usual electrical environment like household electrical appliance. Laptop computers induce EMF to human body, especially to fetus in pregnant women (7). Electric products and consumptions are increasing rapidly in operating theatre as medical science advances. This leads to more opportunities of exposure to ELF-EMF in operating room. Various system of human body could be influenced by EMF like hematic, neurologic, reproductive, endocrine and genetic structure. ELF-EMF is still controversial that critically dangerous to human. However, there was growing evidence that ELF-EMF is harmful in human and animal research. Animal study can be conducted in high level of EMF, but human study is usually based on epidemiologic data. Acute non-lymphocytic leukemia was significantly increased in electrical workers of New Zealand (8). Children environmentally exposed to high EMF have a risk of high incidence of leukemia. EMF might be carcinogen in childhood leukemia and brain tumor (9). Unexplained spontaneous abortion is probably related to ELF-EMF in Iranian women population (10). Occupational exposure to EMF can alter the serum cortisol level in dentists and dentistry students (11). ELF-EMF has a biological stress and damage on prostate gland of rat during pregnancy and postnatal period in 80 Gauss condition (12). Long-term exposure of EMF can affect bone metabolism, followed by decrease of calcium, zinc and magnesium in rat (13). ELF-EMF has potential risk of genotoxic effect (14). DNA strand can be affected by EMF in high level, 1000 μT (15). Humans are revealed to several kinds of carcinogen simultaneously, and EMF can promote carcinogenesis with other carcinogen (16). EMF has not only hazardous effect, but also positive effect. EMF exposure is followed by anti-inflammatory effect in pathologic condition, especially chronic wound healing. Cytokine is modulated due to ELF-EMF, and contributes bioavailability of nitric oxide (17). Perioperative hypothermia has several detrimental effects as followings: arrhythmia, coagulopathy, decreased metabolism, wound infection and mental change (18). Anesthesiologists make an effort to maintain body temperature using various methods like airway heating, warm intravenous fluid, circulating water mattress and forced air (19). Convective air warming system is effective to prevent hypothermia combined with warming blanket. There are several kinds of blanket offered by a manufacturing company for upper body, lower body and pediatric use. Forced air warmer (WarmTouchTM) generates ELF-EMF like other electronic manufactures. ELF-EMF is inversely proportional to square distance due to the nature of physics. In this study, intensity of ELF-EMF is almost a quarter in twofold distances (Fig. 3). Electricity is the factor that primarily affects ELF-EMF value. High temperature set has a possibility of strong ELF-EMF because of more electric consumption. High, medium, low and ambient temperature set demonstrate intensity of ELF-EMF in order (Fig. 2). ELF-EMF should be considered in occupational health protection. More electromagnetic radiation leads to abnormal value of cardiovascular, liver and hematology system in Chinese workers (20, 21). Safe standard of ELF-EMF was under 2 mG recommended by TCO guideline, Swedish Confederation of Professional Employees in the distance of 30cm (22). To protect workers, it is necessary to put distance from EMF source or physical protectors between them.

Conclusion

ELF-EMF from convective air warming system is possibly harmful in close distance. Therefore, medical personnel and patients are safe to establish sufficient distance from ELF-EMF source.

Ethical considerations

Ethical issues (Including plagiarism, informed consent, misconduct, data fabrication and/or falsification, double publication and/or submission, redundancy, etc.) have been completely observed by the authors.
  18 in total

Review 1.  Non-ionizing radiation, Part 1: static and extremely low-frequency (ELF) electric and magnetic fields.

Authors: 
Journal:  IARC Monogr Eval Carcinog Risks Hum       Date:  2002

2.  Residential and occupational exposure to 50-Hz magnetic fields and brain tumours in Norway: a population-based study.

Authors:  Lars Klaeboe; Karl Gerhard Blaasaas; Tor Haldorsen; Tore Tynes
Journal:  Int J Cancer       Date:  2005-05-20       Impact factor: 7.396

3.  Exposure to electromagnetic fields from laptop use of "laptop" computers.

Authors:  C V Bellieni; I Pinto; A Bogi; N Zoppetti; D Andreuccetti; G Buonocore
Journal:  Arch Environ Occup Health       Date:  2012       Impact factor: 1.663

4.  Extremely low-frequency magnetic field decreased calcium, zinc and magnesium levels in costa of rat.

Authors:  Refik Ulku; Mehmet Zulkuf Akdag; Sait Erdogan; Zeki Akkus; Suleyman Dasdag
Journal:  Biol Trace Elem Res       Date:  2010-09-25       Impact factor: 3.738

Review 5.  Extremely low frequency electromagnetic field and wound healing: implication of cytokines as biological mediators.

Authors:  Mirko Pesce; Antonia Patruno; Lorenza Speranza; Marcella Reale
Journal:  Eur Cytokine Netw       Date:  2013-03       Impact factor: 2.737

6.  [Effect of exposure to extremely low-frequency electromagnetic fields on liver function of workers].

Authors:  Xin Liu; Long-yu Zhao; Hui-ling Chen; Cong Liu; Xiao-dong Liu; Shu-mei Ma
Journal:  Zhonghua Lao Dong Wei Sheng Zhi Ye Bing Za Zhi       Date:  2013-08

7.  Exposure to extremely low frequency electromagnetic fields during pregnancy and the risk of spontaneous abortion: a case-control study.

Authors:  Fatemeh Shamsi Mahmoudabadi; Saeideh Ziaei; Mohammad Firoozabadi; Anoshirvan Kazemnejad
Journal:  J Res Health Sci       Date:  2013-09-17

8.  Acute leukemia in electrical workers: a New Zealand case-control study.

Authors:  P Bethwaite; A Cook; J Kennedy; N Pearce
Journal:  Cancer Causes Control       Date:  2001-10       Impact factor: 2.506

9.  Induction of DNA strand breaks by intermittent exposure to extremely-low-frequency electromagnetic fields in human diploid fibroblasts.

Authors:  Sabine Ivancsits; Elisabeth Diem; Alexander Pilger; Hugo W Rüdiger; Oswald Jahn
Journal:  Mutat Res       Date:  2002-08-26       Impact factor: 2.433

Review 10.  Do electromagnetic fields enhance the effects of environmental carcinogens?

Authors:  Jukka Juutilainen
Journal:  Radiat Prot Dosimetry       Date:  2008-10-30       Impact factor: 0.972

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.