Evaluation of 16 New Holmium:Yttrium-Aluminum-Garnet Laser Optical Fibers for Ureteroscopy.

Erin C Akar1, Bodo E Knudsen2.   

Abstract

OBJECTIVE: To test the performance of 16 new single-use holmium:yttrium-aluminum-garnet (YAG) laser fibers.
MATERIALS AND METHODS: Small and medium core fibers were evaluated for flexibility, true diameter, connector temperature, and failure threshold. A flexible ureteroscope was deflected with the fiber in the working channel to measure flexibility. Diameter was measured by micrometer and connector temperature by infrared thermometer. Failure threshold was determined by bending the fiber to 180°, beginning with a radius of 1.25 cm. A 100 W holmium:YAG laser was operated at 1.2 J/10 Hz for 30 seconds or until fiber fracture. The radius was decreased in 0.25-cm increments until a minimum bend radius of 0.4 cm was attained or until fiber fracture.
RESULTS: Of the small core-fibers, the Cook-HLF-S150 (Cook Medical) had the smallest diameter and the Flexiva TracTip 200 (Boston Scientific) the largest. The Cook-HLF-S150 and S200 were the most flexible and the SlimLine EZ200 (Lumenis) the least. The SlimLine EZ200 failed at the largest bend radius, whereas the Flexiva 200 and Flexiva TracTip 200 did not fracture. Of the medium-core fibers, the ScopeSafe 300 had the smallest diameter and the Flexiva 365 the largest. The ScopeSafe 300 was the most flexible and the SlimLine 365 the least. The ScopeSafe 365 failed at the largest radius of 1.25 cm, and the Flexiva 365 did not fail in 6 of 9 trials at the tightest radius.
CONCLUSION: Performance characteristics of these new holmium:YAG optical fibers differed significantly but performance was on par or better than historical controls.
Copyright © 2015 Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2015        PMID: 26169001     DOI: 10.1016/j.urology.2015.04.017

Source DB:  PubMed          Journal:  Urology        ISSN: 0090-4295            Impact factor:   2.649


  4 in total

1.  Impact of laser fiber tip cleavage on power output for ureteroscopy and stone treatment.

Authors:  M Haddad; E Emiliani; Y Rouchausse; F Coste; L Berthe; S Doizi; S Buttice; B Somani; O Traxer
Journal:  World J Urol       Date:  2017-05-30       Impact factor: 4.226

Review 2.  Intracorporeal lithotripsy.

Authors:  Peter Alken
Journal:  Urolithiasis       Date:  2017-12-04       Impact factor: 3.436

Review 3.  Endoscopic management of upper tract urothelial carcinoma-tips and tricks.

Authors:  Asaf Shvero; Dorit E Zilberman; Zohar A Dotan; Maneham Laufer; Eddie Fridman; Harry Winkler; Nir Kleinmann
Journal:  Transl Androl Urol       Date:  2020-08

Review 4.  Advances in Lasers for the Treatment of Stones-a Systematic Review.

Authors:  Peter Kronenberg; Bhaskar Somani
Journal:  Curr Urol Rep       Date:  2018-05-17       Impact factor: 3.092

  4 in total