Literature DB >> 26161111

Patient-prosthesis mismatch after mitral valve-in-valve procedure - at the cost of life or serious consequence?

Dariusz Jagielak1, Radoslaw Kozaryn1, Miłosz Jaguszewski2, Jan Rogowski1.   

Abstract

Entities:  

Year:  2015        PMID: 26161111      PMCID: PMC4495135          DOI: 10.5114/pwki.2015.52292

Source DB:  PubMed          Journal:  Postepy Kardiol Interwencyjnej        ISSN: 1734-9338            Impact factor:   1.426


× No keyword cloud information.
An 81-year-old man, who had undergone a complex cardiosurgical procedure 8 years previously (aortic valve replacement by Medtronic Hancock II 23 mm and mitral valve replacement by Medtronic Hancock II 31 mm, tricuspid valve repair using the Kay technique, and coronary artery bypass surgery with left internal mammary artery (LIMA) to left anterior descending (LAD)), was admitted to the hospital with symptoms suggestive of acute heart failure (NYHA IV). Echocardiographic examination revealed satisfactory function of the aortic prosthesis, and massive regurgitation of the degenerative mitral prosthesis. Left ventricle ejection fraction was estimated at approximately 30% with left ventricle (LV) thinning, an akinetic and thin apex, and mild tricuspid valve insufficiency. Due to these findings cardiopulmonary bypass was used. An intra-aortic balloon pump (IABP) was not introduced due to limited femoral access. The patient, in spite of a very high operative risk (STS 29.8%), was referred for urgent redo cardiac surgery. With this in mind, the patient was qualified for transcatheter mitral valve-in-valve (VIV) implantation. The procedure was performed on-pump with the cardiopulmonary bypass connected via femoral vessels. A transapical approach for delivering the 29 mm Edwards Sapien valve into the mitral position was chosen. Significant efforts were made to protect the scarred apex (pericardium was not dissected). Intraoperative transesophageal echocardiogram (TEE) and fluoroscopy demonstrated precise landing of the new prosthesis, as well as resolution of prior regurgitation and absence of perivalvular leakage. Unfortunately, the initially high mean gradient through the valve remained elevated. The postoperative course was uneventful. On postoperative day 5 in the Cardiology Department due to left ventricular end-diastolic diameter (LVED) > 70 mm and one incidence of tachyarrhythmia, a cardioverter-defibrillator was implanted prophylactically. The patient was discharged home 2 weeks after the procedure. The patient remains in stable condition in class II/III NYHA after 10 months of follow-up. Control echocardiography showed moderate/severe stenosis (mitral valve area 1.0 cm2, mean diastolic pressure gradient (PG) 9 mm Hg) of the implanted VIV prosthesis with no perivalvular leakage and degenerative changes of the leaflets (Figure 1).
Figure 1

A, B – Angio-CT – internal diameter sizing of the bioprosthesis annulus (A – 2.58 cm, B – 2.78 cm). C – Doppler showing no regurgitation, but the orifice area is still not optimal. D – Angiography after implantation

A, B – Angio-CT – internal diameter sizing of the bioprosthesis annulus (A – 2.58 cm, B – 2.78 cm). C – Doppler showing no regurgitation, but the orifice area is still not optimal. D – Angiography after implantation Though the problem of low mitral valve orifice after transcatheter VIV implantation has been deliberated in the literature, in our opinion it has not been outlined in sufficient detail [1]. It is not entirely clear if the patient-prosthesis mismatch (PPM) in our case was influenced by the anatomy of this particular patient, or is rather caused by the atypical position of the bioprosthesis, which was originally designed for the aortic position [2]. From a technical point of view the procedure was executed correctly with maximum balloon inflation during implantation and precise prosthesis positioning. The range of mean diastolic pressure gradients through the mitral orifice after VIV procedures is expansive in available data. We presume, according to our experience, that the rigidity of the biological tissue of which the leaflets are made does not allow complete opening under the diastolic pressure gradients between the left atrium (LA) and LV, which are physiologically much lower than that of the aortic valve [3]. Although VIV has been performed approximately a few hundred times worldwide, there are still quite limited data regarding follow-up of these patients after this procedure [4].
  4 in total

1.  Valve-in-valve procedure: importance of the anatomy of surgical bioprostheses.

Authors:  Alia Noorani; Rizwan Attia; Vinayak Bapat
Journal:  Multimed Man Cardiothorac Surg       Date:  2014-10-29

2.  Neo-annulus: a reference plane in a surgical heart valve to facilitate a valve-in-valve procedure.

Authors:  Vinayak Bapat; Benjamin Adams; Rizwan Attia; Alia Noorani; Martyn Thomas
Journal:  Catheter Cardiovasc Interv       Date:  2014-07-25       Impact factor: 2.692

3.  Use of transcatheter heart valves for a valve-in-valve implantation in patients with degenerated aortic bioprosthesis: technical considerations and results.

Authors:  Vinayak Bapat; Rizwan Attia; Simon Redwood; Jane Hancock; Karen Wilson; Christopher Young; Martyn Thomas
Journal:  J Thorac Cardiovasc Surg       Date:  2012-12       Impact factor: 5.209

Review 4.  Transcatheter valve-in-valve implantation: a systematic review of literature.

Authors:  Jwalant Raval; Vinayak Nagaraja; Guy D Eslick; A Robert Denniss
Journal:  Heart Lung Circ       Date:  2014-06-24       Impact factor: 2.975

  4 in total
  1 in total

1.  Transcatheter transapical valve-in-valve implantation for degenerated mitral bioprosthesis.

Authors:  Michał Lelek; Wojciech Wojakowski; Damian Hudziak; Anna Rybicka-Musialik; Katarzyna Mizia-Stec
Journal:  Postepy Kardiol Interwencyjnej       Date:  2017-11-29       Impact factor: 1.426

  1 in total