| Literature DB >> 26158075 |
Niti M Dhutia1, Graham D Cole1, Massoud Zolgharni1, Charlotte H Manisty1, Keith Willson1, Kim H Parker2, Alun D Hughes1, Darrel P Francis1.
Abstract
Obtaining a "correct" view in echocardiography is a subjective process in which an operator attempts to obtain images conforming to consensus standard views. Real-time objective quantification of image alignment may assist less experienced operators, but no reliable index yet exists. We present a fully automated algorithm for detecting incorrect medial/lateral translation of an ultrasound probe by image analysis. The ability of the algorithm to distinguish optimal from sub-optimal four-chamber images was compared to that of specialists-the current "gold-standard." The orientation assessments produced by the automated algorithm correlated well with consensus visual assessments of the specialists ([Formula: see text]) and compared favourably with the correlation between individual specialists and the consensus, [Formula: see text]. Each individual specialist's assessments were within the consensus of other specialists, [Formula: see text] of the time, and the algorithm's assessments were within the consensus of specialists 85% of the time. The mean discrepancy in probe translation values between individual specialists and their consensus was [Formula: see text], and between the automated algorithm and specialists' consensus was [Formula: see text]. This technology could be incorporated into hardware to provide real-time guidance for image optimisation-a potentially valuable tool both for training and quality control.Entities:
Keywords: automated guidance; echocardiography; speckle tracking; ultrasound probe positioning
Year: 2014 PMID: 26158075 PMCID: PMC4478759 DOI: 10.1117/1.JMI.1.3.037001
Source DB: PubMed Journal: J Med Imaging (Bellingham) ISSN: 2329-4302