S J HosseiniAliabadi1, S M Hosseini Pooya2, H Afarideh3, F Mianji2.
Abstract
INTRODUCTION: The angular dependency of response for TLD cards may cause deviation from its true value on the results of environmental dosimetry, since TLDs may be exposed to radiation at different angles of incidence from the surrounding area.Entities:
Keywords: Accuracy; Calibration; Dosimetry; Environmental; TLD
Year: 2015 PMID: 26157729 PMCID: PMC4479385
Source DB: PubMed Journal: J Biomed Phys Eng ISSN: 2251-7200
Figure 1Schematic of a 3D TL Dosimeter in a Cylindrical Al Wall for Environmental Measurements
Figure 2Schematic of a Three-dimensional Calibration Method with Ten Angles of Incidence
Figure 3Configuration of an Experimental Test of the 3D Dosimeter Response in a Real Radiation Field
Results of Calibration Factors of TLD Cards in 1D and 3D Calibration Methods
| Calibration Method | Card ID #1 | Card ID #2 | Card ID #3 |
|---|---|---|---|
| 1D Calibration Factor of Cards | 0.119 | 0.117 | 0.116 |
| 3D Calibration Factor of Cards | 0.130 | 0.125 | 0.123 |
| 3D Calibration Factor of the 3D Dosimeter System | 0.126 | ||
The Results of Absolute Deviations of TLD Responses Using 1D and 3D Calibration Factors
|
|
|
|
|
|
|
|
|---|---|---|---|---|---|---|
| #1 | 3.41 | 0.43 | 11.7% | 3.64 | 0.20 | 5.2% |
| #2 | 3.13 | 0.71 | ||||
| #3 | 3.64 | 0.20 |