Bipolar radiofrequency ablation with 2 × 2 electrodes as a building block for matrix radiofrequency ablation: Ex vivo liver experiments and finite element method modelling.

Stefaan Mulier1,2, Yansheng Jiang2, Jacques Jamart3, Chong Wang4, Yuanbo Feng2, Guy Marchal2, Luc Michel5, Yicheng Ni2.   

Abstract

PURPOSE: Size and geometry of the ablation zone obtained by currently available radiofrequency (RF) electrodes is highly variable. Reliability might be improved by matrix radiofrequency ablation (MRFA), in which the whole tumour volume is contained within a cage of x × y parallel electrodes. The aim of this study was to optimise the smallest building block for matrix radiofrequency ablation: a recently developed bipolar 2 × 2 electrode system.
MATERIALS AND METHODS: In ex vivo bovine liver, the parameters of the experimental set-up were changed one by one. In a second step, a finite element method (FEM) modelling of the experiment was performed to better understand the experimental findings.
RESULTS: The optimal power to obtain complete ablation in the shortest time was 50-60 W. Performing an ablation until impedance rise was superior to ablation for a fixed duration. Increasing electrode diameter improved completeness of ablation due to lower temperature along the electrodes. A chessboard pattern of electrode polarity was inferior to a row pattern due to an electric field void in between the electrodes. Variability of ablation size was limited. The FEM correctly simulated and explained the findings in ex vivo liver.
CONCLUSIONS: These experiments and FEM modelling allowed a better insight in the factors influencing the ablation zone in a bipolar 2 × 2 electrode RF system. With optimal parameters, complete ablation was obtained quickly and with limited variability. This knowledge will be useful to build a larger system with x × y electrodes for MRFA.

Entities:  

Keywords:  Bipolar; finite element method; liver; matrix; radiofrequency ablation

Mesh:

Year:  2015        PMID: 26156212     DOI: 10.3109/02656736.2015.1046194

Source DB:  PubMed          Journal:  Int J Hyperthermia        ISSN: 0265-6736            Impact factor:   3.914


  1 in total

1.  A piecewise function of resistivity of liver: determining parameters with finite element analysis of radiofrequency ablation.

Authors:  Ricardo Possebon; Yansheng Jiang; Stefaan Mulier; Chong Wang; Feng Chen; Yuanbo Feng; Yicheng Ni
Journal:  Med Biol Eng Comput       Date:  2017-08-02       Impact factor: 2.602

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.