The SOMNOtouch device as a novel method for measuring short-term blood pressure variability: a comparison with the Finometer.

Lisa S Manning1, Thompson G Robinson, Ronney B Panerai.   

Abstract

BACKGROUND: Current noninvasive techniques to capture short-term blood pressure variability (BPV) have methodological and practical limitations. This study assessed the ability of a novel device, the SOMNOtouch, which derives continuous blood pressure (BP) measures from pulse transit time, to estimate BPV, compared with the widely used Finometer.
METHODS: BP monitoring was performed simultaneously on the SOMNOtouch and Finometer devices in 16 healthy volunteers. Systolic and diastolic BPVs, defined as SD and coefficient of variation, were derived from measurements from each device for three predefined periods: 0-3, 7-10, and 0-10 min.
RESULTS: Agreement in BPV indices from the two devices was assessed using the Bland-Altman technique. For all BPV parameters, over all measurement periods, broad scatter was observed on Bland-Altman plots. Bias (limits of agreement) for minutes 0-10: SD of systolic BP, -3.03 mmHg (-10.88 to +4.55), SD of diastolic BP -1.65 mmHg (-4.41 to +1.11).
CONCLUSIONS: The poor agreement observed in BPV estimates between the devices may reflect the inability of the current pulse transit time method to sensitively detect changes in BP. Further investigation is needed before such methods can be reliably used to measure short-term BPV.

Mesh:

Year:  2015        PMID: 26154853     DOI: 10.1097/MBP.0000000000000128

Source DB:  PubMed          Journal:  Blood Press Monit        ISSN: 1359-5237            Impact factor:   1.444


  2 in total

1.  Quantitative Assessment of the Impact of Blood Pulsation on Intraocular Pressure Measurement Results in Healthy Subjects.

Authors:  Robert Koprowski; Lei Tian
Journal:  J Ophthalmol       Date:  2017-01-30       Impact factor: 1.909

2.  Feasibility of a new free mobility procedure to evaluate the function of the autonomic nervous system in patients with syncope.

Authors:  Juan Nader-Kawachi; Paulo C Manrique-Mirón; Yaima C Pino-Peña; María L Andrade-Magdaleno; Jesús López-Estrada
Journal:  Sci Rep       Date:  2020-08-19       Impact factor: 4.379

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.