Literature DB >> 26152685

Optimization of stent implantation using a high pressure inflation protocol.

Srikanth Vallurupalli1, Amit Bahia1, Ernesto Ruiz-Rodriguez1, Zubair Ahmed1,2, Abdul Hakeem1,2, Barry F Uretsky1,2.   

Abstract

BACKGROUND: High-pressure inflation is the universal standard for stent deployment but a specific protocol for its use is lacking. We developed a standardized "pressure optimization protocol" (POP) using time to inflation pressure stability as an endpoint for determining the required duration of stent inflation.
OBJECTIVES: The primary study purpose was to determine the stent inflation time (IT) in a large patient cohort using the standardized inflation protocol, to correlate various patient and lesion characteristics with IT, and ascertain in an in vitro study the time for pressure accommodation within an inflation system.
METHODS: Six hundred fifteen stent implants in 435 patients were studied. Multivariate analysis was performed to determine predictors of longer ITs. In an in vitro study, various stents and balloons were inflated in air to determine the pressure accommodation time of the inflation system.
RESULTS: The mean stent IT was 104 ± 41 sec (range 30-380 sec). Stent length was the only predictor of prolonged stent inflation. The "accommodation time" in vitro of the stent inflation system itself was 33 ± 24 sec. The protocol was safe requiring premature inflation termination in <3% of stent implants. No serious adverse events occurred.
CONCLUSIONS: Achieving stable inflation pressure requires on average over 100 sec and may require several minutes in individual cases. Stent length increases IT. These results suggest that the widespread practice of rapid inflation/deflation may not be sufficient to fully expand the stent and that the use of a pressure stability protocol will allow for safe, predictable, and more complete stent deployment.
© 2015 Wiley Periodicals, Inc. © 2015 Wiley Periodicals, Inc.

Entities:  

Keywords:  angioplasty; coronary stenting; percutaneous coronary intervention; stenting technique

Mesh:

Year:  2015        PMID: 26152685     DOI: 10.1002/ccd.26095

Source DB:  PubMed          Journal:  Catheter Cardiovasc Interv        ISSN: 1522-1946            Impact factor:   2.692


  3 in total

Review 1.  Hemodynamic Support Devices for Shock and High-Risk PCI: When and Which One.

Authors:  George W Vetrovec
Journal:  Curr Cardiol Rep       Date:  2017-08-31       Impact factor: 2.931

2.  A randomized controlled clinical trial of prolonged balloon inflation during stent deployment strategy in primary percutaneous coronary intervention for ST-segment elevation myocardial infarction: a pilot study.

Authors:  Min Ma; Ling Wang; Kai-Yue Diao; Shi-Chu Liang; Ye Zhu; Hua Wang; Mian Wang; Li Zhang; Zhi-Gang Yang; Yong He
Journal:  BMC Cardiovasc Disord       Date:  2022-02-04       Impact factor: 2.298

3.  Guidezilla™ guide extension catheter I for transradial coronary intervention.

Authors:  Xinjun Lei; Qi Liang; Yuan Fang; Yihui Xiao; Dongqi Wang; Maozhi Dong; Jiancheng Li; Ting Yu
Journal:  Front Cardiovasc Med       Date:  2022-08-17
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.