Fiber-optic immunosensor for detection of Crimean-Congo hemorrhagic fever IgG antibodies in patients.

Fairoz Algaar1, Evgeni Eltzov2,3, Marina M Vdovenko4, Ivan Yu Sakharov4, Luka Fajs3, Manfred Weidmann5, Ali Mirazimi6,7, Robert S Marks1,2,3,8,9.   

Abstract

Crimean-Congo hemorrhagic fever (CCHF) is a severe viral disease with high fatality rate. CCHF virus is endemic in parts of Africa, Asia, the Middle East, and southeastern Europe. Rapid diagnostics of CCHF is vital for appropriate clinical management and prevention of secondary spread from human-to-human. Currently, diagnostics relies on real-time RT-PCR and antibody or antigen detection using ELISA. These methods require trained personnel and expensive equipment and are not appropriate for point-of-care (POC) diagnostics. Furthermore, there are no POC assays available for CCHF. We developed a fiber-optic biosensor for the detection of CCHF IgG antibodies. In order to improve sensitivity, we optimized both the bioreceptor immobilization protocol and the chemiluminescence substrate formulation. The resulting protocol showed a 100-fold greater sensitivity for detection of CCHF antibodies. Finally, we evaluated the fiber-optic biosensor with two CCHF patient sera. We showed that the fiber-optic biosensor is 10-times more sensitive than colorimetric ELISA and is able to detect both patients with high and low levels of IgG antibodies. We believe that the fiber-optic biosensor is a suitable alternative to ELISA as it is much more sensitive and makes it possible to detect a small amount of antibodies at an early stage of infection and can be integrated as a point-of-care diagnostic system of CCHF.

Entities:  

Mesh:

Substances:

Year:  2015        PMID: 26151547     DOI: 10.1021/acs.analchem.5b01728

Source DB:  PubMed          Journal:  Anal Chem        ISSN: 0003-2700            Impact factor:   6.986


  6 in total

Review 1.  Construction and Potential Applications of Biosensors for Proteins in Clinical Laboratory Diagnosis.

Authors:  Xuan Liu; Hui Jiang
Journal:  Sensors (Basel)       Date:  2017-12-04       Impact factor: 3.576

Review 2.  Point-of-Care-Testing in Acute Stroke Management: An Unmet Need Ripe for Technological Harvest.

Authors:  Dorin Harpaz; Evgeni Eltzov; Raymond C S Seet; Robert S Marks; Alfred I Y Tok
Journal:  Biosensors (Basel)       Date:  2017-08-03

3.  Capture-Layer Lateral Flow Immunoassay: A New Platform Validated in the Detection and Quantification of Dengue NS1.

Authors:  Tim Axelrod; Evgeni Eltzov; Robert S Marks
Journal:  ACS Omega       Date:  2020-05-04

4.  The Incorporation of Amplified Metal-Enhanced Fluorescence in a CMOS-Based Biosensor Increased the Detection Sensitivity of a DNA Marker of the Pathogenic Fungus Colletotrichum gloeosporioides.

Authors:  Dorin Harpaz; Noam Alkan; Evgeni Eltzov
Journal:  Biosensors (Basel)       Date:  2020-12-13

5.  Exploring Sensitive Label-Free Multiplex Analysis with Raman-Coded Microbeads and SERS-Coded Reporters.

Authors:  Umar Azhar; Qazi Ahmed; Saira Ishaq; Zeyad T Alwahabi; Sheng Dai
Journal:  Biosensors (Basel)       Date:  2022-02-16

Review 6.  Immunosensing Based on Optical Fiber Technology: Recent Advances.

Authors:  Maria Simone Soares; Miguel Vidal; Nuno F Santos; Florinda M Costa; Carlos Marques; Sónia O Pereira; Cátia Leitão
Journal:  Biosensors (Basel)       Date:  2021-08-30
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.