Sridhar Sahoo1, Selvam T Palani1, S K Saxena2, D A R Babu1, A Dash2.
Abstract
Skin cancer treatment involving (32)P source is an easy, less expensive method of treatment limited to small and superficial lesions of approximately 1 mm deep. Bhabha Atomic Research Centre (BARC) has indigenously developed (32)P nafion-based patch source (1 cm × 1 cm) for treating skin cancer. For this source, the values of dose per unit activity at different depths including dose profiles in water are calculated using the EGSnrc-based Monte Carlo code system. For an initial activity of 1 Bq distributed in 1 cm(2) surface area of the source, the calculated central axis depth dose values are 3.62 × 10(-10) GyBq(-1) and 8.41 × 10(-11) GyBq(-1)at 0.0125 and 1 mm depths in water, respectively. Hence, the treatment time calculated for delivering therapeutic dose of 30 Gy at 1 mm depth along the central axis of the source involving 37 MBq activity is about 2.7 hrs.Entities:
Keywords: 32P patch; brachytherapy; dosimetry; nafion; skin cancer
Year: 2015 PMID: 26150682 PMCID: PMC4471639 DOI: 10.4103/0971-6203.152232
Source DB: PubMed Journal: J Med Phys ISSN: 0971-6203
Figure 1(a) Schematic diagram of the 32P-nafion-patch source and water phantom used in the DOSXYZnrc Monte Carlo simulation. (b) Coordinate system used in the simulation
Elemental composition of phosphorous-loaded zirconium-nafion- 117 composite membrane (density= 1620 kgm−3)
Figure 232P beta spectrum used in the Monte Carlo simulation
Figure 3Depth dose distribution of 32P-nafion-patch source along the central axis of the source
Comparison of dose values per unit activity (GyBq−1) presented as a function of depth in water
Figure 4Dose profile along the x-axis of 32P-nafion-patch source for different depths, z = 0.5 mm, z = 1 mm and z = 2 mm
Figure 5Normalized dose profile along the x-axis of 32P-nafion-patch source for depth z = 1 mm
Figure 6Isodose profiles of the 32P-nafion-patch source on xy-plane at a depth of 0.5 mm
Figure 8Isodose profiles of the 32P-nafion-patch source on xy-plane at a depth of 2 mm