Quality and dose optimization in hand computed radiography.

A L M Pavan1, A F F Alves2, S B Duarte3, G Giacomini4, T Sardenberg5, J R A Miranda6, D R Pina7.   

Abstract

The objective of the present study was to optimize a radiographic technique for hand examinations using a computed radiography (CR) system and demonstrate the potential for dose reductions compared with clinically established technique. An exposure index was generated from the optimized technique to guide operators when imaging hands. Homogeneous and anthropomorphic phantoms that simulated a patient's hand were imaged using a CR system at various tube voltages and current settings (40-55 kVp, 1.25-2.8 mAs), including those used in clinical routines (50 kVp, 2.0 mAs) to obtain an optimized chart. The homogeneous phantom was used to assess objective parameters that are associated with image quality, including the signal difference-to-noise ratio (SdNR), which is used to define a figure of merit (FOM) in the optimization process. The anthropomorphic phantom was used to subjectively evaluate image quality using Visual Grading Analysis (VGA) that was performed by three experienced radiologists. The technique that had the best VGA score and highest FOM was considered the gold standard (GS) in the present study. Image quality, dose and the exposure index that are currently used in the clinical routine for hand examinations in our institution were compared with the GS technique. The effective dose reduction was 67.0%. Good image quality was obtained for both techniques, although the exposure indices were 1.60 and 2.39 for the GS and clinical routine, respectively.
Copyright © 2015 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.

Entities:  

Keywords:  Figure of merit; Image optimization; Phantom; Visual Grading Analysis

Mesh:

Year:  2015        PMID: 26148866     DOI: 10.1016/j.ejmp.2015.06.010

Source DB:  PubMed          Journal:  Phys Med        ISSN: 1120-1797            Impact factor:   2.685


  1 in total

1.  Homogeneous Canine Chest Phantom Construction: A Tool for Image Quality Optimization.

Authors:  Ana Luiza Menegatti Pavan; Maria Eugênia Dela Rosa; Guilherme Giacomini; Fernando Antonio Bacchim Neto; Seizo Yamashita; Luiz Carlos Vulcano; Sergio Barbosa Duarte; José Ricardo de Arruda Miranda; Diana Rodrigues de Pina
Journal:  PLoS One       Date:  2016-04-21       Impact factor: 3.240

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.