How physical infusion system parameters cause clinically relevant dose deviations after setpoint changes.

Annemoon M Timmerman, Roland A Snijder, Peter Lucas, Martine C Lagerweij, Joris H Radermacher, Maurits K Konings.   

Abstract

Multi-infusion therapy, in which multiple pumps are connected to one access point, is frequently used in patient treatments. This practice is known to cause dosing errors following setpoint changes in the drug concentrations that actually enter the patients. Within the Metrology for Drug Delivery Project, we analyzed and quantified the two main physical phenomena leading to these errors: the "push-out" effect and the system mechanical compliance. We compared the dosing errors of a three-pump system with two infusion sets, both with and without anti-reflux valves, using in vitro spectrophotometric experiments. Additionally, computer simulations were used to study the compliance effect separately. We found a start-up time of more than 1 h, and a dosing error following a setpoint increase of another pump for the low flow rate pump, corresponding to 0.5 μg noradrenaline delivered in 8 min. We showed that the dead volume inside the tubes and syringe compliance produce opposite deviations from the setpoint values in the actual drug output concentrations, making the net result hard to predict and often counterintuitive. We conclude that metrology on compliance and push-out effects could be used by infusion device manufacturers to successfully improve drug delivery performance and relevant standards for high-risk multi-infusion applications.

Entities:  

Mesh:

Year:  2015        PMID: 26146091     DOI: 10.1515/bmt-2014-0139

Source DB:  PubMed          Journal:  Biomed Tech (Berl)        ISSN: 0013-5585            Impact factor:   1.411


  2 in total

1.  Analytical method for calculation of deviations from intended dosages during multi-infusion.

Authors:  Maurits K Konings; Roland A Snijder; Joris H Radermacher; Annemoon M Timmerman
Journal:  Biomed Eng Online       Date:  2017-01-17       Impact factor: 2.819

2.  Disturbance of Vancomycin Infusion Flow during Multidrug Infusion: Influence on Endothelial Cell Toxicity.

Authors:  Maryline Drouet; Elodie Cuvelier; Feng Chai; Stéphanie Genay; Pascal Odou; Bertrand Décaudin
Journal:  Antibiotics (Basel)       Date:  2021-12-24
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.