Novel Navigated Ultrasound Compared With Conventional Ultrasound for Vascular Access—a Prospective Study in a Gel Phantom Model.

David Sander1, Volker Schick2, Hannes Ecker2, Falko Lindacher2, Moritz Felsch3, Oliver Spelten2, Robert Schier2, Jochen Hinkelbein2, Stephan A Padosch2.   

Abstract

OBJECTIVES: The authors hypothesized that, compared with conventional ultrasound (CUS), the use of a novel navigated ultrasound (NUS) technology would increase success rates and decrease performance times of vascular access procedures in a gel phantom model.
DESIGN: A prospective, randomized, crossover study.
SETTING: A university Hospital. PARTICIPANTS: Participants were 44 anesthesiologists with varying clinical experience.
INTERVENTIONS: Anesthesiologists performed in-plane and out-of-plane vascular access procedures using both NUS and CUS for needle visualization in a gel phantom model.
MEASUREMENTS AND MAIN RESULTS: Procedure time was measured from needle insertion to verbalization of final needle positioning by the participants, and successful needle placement into the simulated vessel was verified by aspiration of simulated blood. By employing ultrasound navigation capabilities in addition to real-time ultrasound imaging during in-plane/long-axis vascular access procedures, median procedure time showed a nonsignificant decrease (7.5 seconds v 13.0 seconds; p = 0.028), and the observed increase in procedure success rate (90.9% v 100%; p = 0.125) did not reach statistical significance. For out-of-plane/short-axis vascular access procedures, a significant reduction in median procedure time (5.0 seconds v 11.5 seconds; p<0.001) and a significant increase in procedure success rate (75% v 100%; p<0.001) were achieved by using navigation technology combined with real-time ultrasound.
CONCLUSIONS: NUS technology improved the performance times and success rates of vascular access procedures conducted by anesthesiologists in a gel phantom model.
Copyright © 2015 Elsevier Inc. All rights reserved.

Entities:  

Keywords:  SonixGPS; Ultrasonix GPS; central catheter; central venous; gel phantom model; real-time ultrasound navigation; ultrasound; vascular access

Mesh:

Substances:

Year:  2015        PMID: 26142368     DOI: 10.1053/j.jvca.2015.03.014

Source DB:  PubMed          Journal:  J Cardiothorac Vasc Anesth        ISSN: 1053-0770            Impact factor:   2.628


  3 in total

1.  Development and Characterization of an Ex Vivo Testing Platform for Evaluating Automated Central Vascular Access Device Performance.

Authors:  Emily N Boice; David Berard; Sofia I Hernandez Torres; Guy Avital; Eric J Snider
Journal:  J Pers Med       Date:  2022-08-05

2.  Computer-assisted needle navigation for pediatric internal jugular central venous cannulation: A feasibility study.

Authors:  Christopher Gallo; Pezhman Foroughi; Elisabeth Meagher; Ranjith Vellody; Bhupender Yadav; Anthony Ho; Alican Demir; Dorothee Heisenberg; Kevin Cleary; Karun Sharma
Journal:  J Vasc Access       Date:  2020-04-27       Impact factor: 2.283

3.  Ultrasound-guided lumbar puncture with a needle-guidance system: A prospective and controlled study to evaluate the learnability and feasibility of a newly developed approach.

Authors:  Tilo Backhaus; Moritz von Cranach; Jochen Brich
Journal:  PLoS One       Date:  2018-04-09       Impact factor: 3.240

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.