I S Singh1, Lokpati Mishra2, J R Yadav2, M Y Nadar2, D D Rao2, K S Pradeepkumar2.
Abstract
The estimation of Pu/(241)Am ratio in the biological samples is an important input for the assessment of internal dose received by the workers. The radiochemical separation of Pu isotopes and (241)Am in a sample followed by alpha spectrometry is a widely used technique for the determination of Pu/(241)Am ratio. However, this method is time consuming and many times quick estimation is required. In this work, Pu/(241)Am ratio in the biological sample was estimated with HPGe detector based measurements using gamma/X-rays emitted by these radionuclides. These results were compared with those obtained from alpha spectroscopy of sample after radiochemical analysis and found to be in good agreement.Entities:
Keywords: Americium; Biological sample; Gamma ray spectroscopy; HPGe detector; Plutonium; Pu/Am ratio; α-Spectroscopy
Mesh:
Substances:
Year: 2015 PMID: 26141295 DOI: 10.1016/j.apradiso.2015.06.012