The Effect of Direct Current Stimulation versus T-Wave Shock on Defibrillation Threshold Testing.

Hiroyuki Ito1, Mitsuharu Kawamura2, Nitish Badhwar1, Vasanth Vedantham1, Zian H Tseng1, Byron K Lee1, Randall J Lee1, Gregory M Marcus1, Edward P Gerstenfeld1, Melvin M Scheinman1.   

Abstract

INTRODUCTION: There are several methods to induce ventricular fibrillation (VF) during defibrillation threshold (DFT) testing. Delivering a shock at a critical time during the T wave (T-shock) is the conventional approach, while delivering a constant direct current voltage (DC stim) from the implantable cardioverter defibrillator is an alternative method. Only a few reports compare VF induction methods. The purpose of this study was to evaluate the effects and safety of DC stim versus T-shock.
METHODS: We retrospectively investigated 414 consecutive patients undergoing DFT testing. We compared the two groups (DC stim and T-shock) with respect to clinical characteristics, electrocardiogram (ECG) changes, and complications.
RESULTS: Ventricular arrhythmia, including ventricular tachycardia (VT) and VF, was induced by DC stim in 93 patients or T-shock in 321 patients. No more than three attempts were performed during one procedure. There was no significant difference in the baseline ECG, induced tachycardia cycle length (TCL), or complications between the two groups. However, the induced TCL was significantly shorter than the clinical TCL regardless of induction method (P = 0.001). Five patients suffered major complications (i.e., electromechanical dissociation or incessant VT). A history of atrial fibrillation was significantly greater in patients with major complications than the others (80% vs 24%, P = 0.004), and was an independent predictor on multivariate analysis.
CONCLUSIONS: There is no significant difference in induced TCL or complications between the DC stim and T-shock. The induced TCL is significantly shorter than clinical TCL regardless of induction method.
© 2015 Wiley Periodicals, Inc.

Entities:  

Keywords:  defibrillation threshold testing; direct current voltage; electromechanical dissociation; shock on T; tachycardia cycle length

Mesh:

Year:  2015        PMID: 26137999     DOI: 10.1111/pace.12684

Source DB:  PubMed          Journal:  Pacing Clin Electrophysiol        ISSN: 0147-8389            Impact factor:   1.976


  1 in total

1.  Acute Effects of Implantable Cardioverter-Defibrillator Shocks on Biomarkers of Myocardial Injury, Apoptosis, Heart Failure, and Systemic Inflammation.

Authors:  Jordan Brewster; Travis Sexton; Gary Dhaliwal; Richard Charnigo; Gustavo Morales; Kevin Parrott; Yousef Darrat; John Gurley; Susan Smyth; Claude S Elayi
Journal:  Pacing Clin Electrophysiol       Date:  2017-03-08       Impact factor: 1.976

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.