Andrea Russo1, Francesco Morescalchi1, Ciro Costagliola2, Luisa Delcassi1, Francesco Semeraro1.
Abstract
Purpose. To construct an inexpensive, convenient, and portable attachment for smartphones for the acquisition of still and live retinal images. Methods. A small optical device based on the principle of direct ophthalmoscopy was designed to be magnetically attached to a smartphone. Representative images of normal and pathological fundi were taken with the device. Results. A field-of-view up to ~20° was captured at a clinical resolution for each fundus image. The cross-polarization technique adopted in the optical design dramatically diminished corneal Purkinje reflections, making it possible to screen patients even through undilated pupils. Light emission proved to be well within safety limits. Conclusions. This optical attachment is a promising, inexpensive, and valuable alternative to the direct ophthalmoscope, potentially eliminating problems of poor exam skills and inexperienced observer bias. Its portability, together with the wireless connectivity of smartphones, presents a promising platform for screening and telemedicine in nonhospital settings. Translational Relevance. Smartphones have the potential to acquire retinal imaging for a portable ophthalmoscopy.Entities:
Year: 2015 PMID: 26137320 PMCID: PMC4468345 DOI: 10.1155/2015/823139
Source DB: PubMed Journal: J Ophthalmol ISSN: 2090-004X Impact factor: 1.909
Figure 1Exploded view of the D-Eye module (angles and distances between components are approximated). Retinal images are acquired using coaxial illumination and imaging paths thanks to a beam splitter (C). The blue arrow depicts the path of the light; red arrow depicts the path of fundus imaging. Device components are glass platelet (A) with imprinted negative lens (A′), photo-absorbing wall (B), beam splitter (C), mirror (D), plastic case (E), diaphragm (F), polarized filters (G, H), flash and camera glass (J, I), and magnetic external ring (K).
Figure 2Picture of the prototypes magnetically attached to different iPhone (5 and 5s; Apple, Cupertino, CA) models.
Figure 4Representative retinal images taken with D-Eye. (a) Normal optic disc in an undilated child. (b) Normal posterior pole in a dilated 29-year-old woman. (c) Dry age-related maculopathy in an undilated 75-year-old man. (d) Optic nerve glioma in a 23-year-old undilated woman. (e) Posterior vitreous detachment in a dilated 72-year-old pseudophakic woman. (f) Waxy disc pallor and pigmentary changes in a 50-year-old man with retinitis pigmentosa. ((g) and (h)) Depiction of the same optic nerve head by D-Eye and Canon CR-2 Retinal Camera.
Figure 5Cross-polarization-accentuated nerve-fiber layer definition (white arrows).