Literature DB >> 26133614

A framework for the correction of slow physiological drifts during MR-guided HIFU therapies: Proof of concept.

Cornel Zachiu1, Baudouin Denis de Senneville2, Chrit Moonen1, Mario Ries1.   

Abstract

PURPOSE: While respiratory motion compensation for magnetic resonance (MR)-guided high intensity focused ultrasound (HIFU) interventions has been extensively studied, the influence of slow physiological motion due to, for example, peristaltic activity, has so far been largely neglected. During lengthy interventions, the magnitude of the latter can exceed acceptable therapeutic margins. The goal of the present study is to exploit the episodic workflow of these therapies to implement a motion correction strategy for slow varying drifts of the target area and organs at risk over the entire duration of the intervention.
METHODS: The therapeutic workflow of a MR-guided HIFU intervention is in practice often episodic: Bursts of energy delivery are interleaved with periods of inactivity, allowing the effects of the beam on healthy tissues to recede and/or during which the plan of the intervention is reoptimized. These periods usually last for at least several minutes. It is at this time scale that organ drifts due to slow physiological motion become significant. In order to capture these drifts, the authors propose the integration of 3D MR scans in the therapy workflow during the inactivity intervals. Displacements were estimated using an optical flow algorithm applied on the 3D acquired images. A preliminary study was conducted on ten healthy volunteers. For each volunteer, 3D MR images of the abdomen were acquired at regular intervals of 10 min over a total duration of 80 min. Motion analysis was restricted to the liver and kidneys. For validating the compatibility of the proposed motion correction strategy with the workflow of a MR-guided HIFU therapy, an in vivo experiment on a porcine liver was conducted. A volumetric HIFU ablation was completed over a time span of 2 h. A 3D image was acquired before the first sonication, as well as after each sonication.
RESULTS: Following the volunteer study, drifts larger than 8 mm for the liver and 5 mm for the kidneys prove that slow physiological motion can exceed acceptable therapeutic margins. In the animal experiment, motion tracking revealed an initial shift of up to 4 mm during the first 10 min and a subsequent continuous shift of ∼2 mm/h until the end of the intervention. This leads to a continuously increasing mismatch of the initial shot planning, the thermal dose measurements, and the true underlying anatomy. The estimated displacements allowed correcting the planned sonication cell cluster positions to the true target position, as well as the thermal dose estimates during the entire intervention and to correct the nonperfused volume measurement. A spatial coherence of all three is particularly important to assure a confluent ablation volume and to prevent remaining islets of viable malignant tissue.
CONCLUSIONS: This study proposes a motion correction strategy for displacements resulting from slowly varying physiological motion that might occur during a MR-guided HIFU intervention. The authors have shown that such drifts can lead to a misalignment between interventional planning, energy delivery, and therapeutic validation. The presented volunteer study and in vivo experiment demonstrate both the relevance of the problem for HIFU therapies and the compatibility of the proposed motion compensation framework with the workflow of a HIFU intervention under clinical conditions.

Entities:  

Mesh:

Year:  2015        PMID: 26133614     DOI: 10.1118/1.4922403

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  10 in total

1.  Elasticity mapping of murine abdominal organs in vivo using harmonic motion imaging (HMI).

Authors:  Thomas Payen; Carmine F Palermo; Stephen A Sastra; Hong Chen; Yang Han; Kenneth P Olive; Elisa E Konofagou
Journal:  Phys Med Biol       Date:  2016-07-12       Impact factor: 3.609

2.  Pancreas deformation in the presence of tumors using feature tracking from free-breathing XD-GRASP MRI.

Authors:  Teodora Chitiboi; Matthew Muckley; Bari Dane; Chenchan Huang; Li Feng; Hersh Chandarana
Journal:  J Magn Reson Imaging       Date:  2019-03-10       Impact factor: 4.813

3.  Motion guided Spatiotemporal Sparsity for high quality 4D-CBCT reconstruction.

Authors:  Yang Liu; Xi Tao; Jianhua Ma; Zhaoying Bian; Dong Zeng; Qianjin Feng; Wufan Chen; Hua Zhang
Journal:  Sci Rep       Date:  2017-12-12       Impact factor: 4.379

4.  Ultrasound image based visual servoing for moving target ablation by high intensity focused ultrasound.

Authors:  Joonho Seo; Norihiro Koizumi; Mamoru Mitsuishi; Naohiko Sugita
Journal:  Int J Med Robot       Date:  2016-12-20       Impact factor: 2.547

5.  Real-time auto-adaptive margin generation for MLC-tracked radiotherapy.

Authors:  M Glitzner; M F Fast; B Denis de Senneville; S Nill; U Oelfke; J J W Lagendijk; B W Raaymakers; S P M Crijns
Journal:  Phys Med Biol       Date:  2016-12-17       Impact factor: 3.609

6.  A framework for continuous target tracking during MR-guided high intensity focused ultrasound thermal ablations in the abdomen.

Authors:  Cornel Zachiu; Baudouin Denis de Senneville; Ivan D Dmitriev; Chrit T W Moonen; Mario Ries
Journal:  J Ther Ultrasound       Date:  2017-10-09

7.  Intrafraction motion during radiotherapy of breast tumor, breast tumor bed, and individual axillary lymph nodes on cine magnetic resonance imaging.

Authors:  Maureen L Groot Koerkamp; H J G Desirée van den Bongard; Marielle E P Philippens; Femke van der Leij; Stefano Mandija; Antonetta C Houweling
Journal:  Phys Imaging Radiat Oncol       Date:  2022-07-05

8.  Real-time 3D motion estimation from undersampled MRI using multi-resolution neural networks.

Authors:  Maarten L Terpstra; Matteo Maspero; Tom Bruijnen; Joost J C Verhoeff; Jan J W Lagendijk; Cornelis A T van den Berg
Journal:  Med Phys       Date:  2021-10-26       Impact factor: 4.506

9.  A hierarchical model of abdominal configuration changes extracted from golden angle radial magnetic resonance imaging.

Authors:  Yuhang Zhang; Rojano Kashani; Yue Cao; Theodore S Lawrence; Adam Johansson; James M Balter
Journal:  Phys Med Biol       Date:  2021-02-09       Impact factor: 3.609

10.  A Nano-Emulsion Platform Functionalized with a Fully Human scFv-Fc Antibody for Atheroma Targeting: Towards a Theranostic Approach to Atherosclerosis.

Authors:  Samuel Bonnet; Geoffrey Prévot; Stéphane Mornet; Marie-Josée Jacobin-Valat; Yannick Mousli; Audrey Hemadou; Mathieu Duttine; Aurélien Trotier; Stéphane Sanchez; Martine Duonor-Cérutti; Sylvie Crauste-Manciet; Gisèle Clofent-Sanchez
Journal:  Int J Mol Sci       Date:  2021-05-14       Impact factor: 5.923

  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.