Literature DB >> 26131025

Deployment Accuracy of the Conformable GORE(®) TAG(®) Thoracic Endoprosthesis in the Treatment of Zones 2 and 3 Aortic Arch Aneurysms Compared with the Previous TAG(®).

Eisaku Ito1, Yuji Kanaoka1, Koji Maeda1, Hiroki Ohta1, Atsushi Ishida1, Takao Ohki1.   

Abstract

PURPOSES: When placing stent grafts, deployment accuracy and birdbeaking due to inadequate conformability and device apposition along the inner curvature of the arch have been encountered. The new Conformable GORE® TAG® Thoracic Endoprosthesis (CTAG) is designed to have enhanced compression resistance and improved conformability in difficult anatomy. The present study compared the deployment accuracy and conformability of the CTAG Device with TAG Device.
METHOD: Deployment accuracy and birdbeaking was compared of CTAG Device and TAG Device implantation for initial treatment of thoracic aortic aneurysm conducted by our department between March 2010 and March 2012. Deployment accuracy was defined as the distance between the actual and intended device implantation locations measured from DSA images.
RESULTS: Deployment accuracy at the time of implantation (mean ± SD) was significantly better for the CTAG Device compared to the TAG Device (2.2 ± 1.7 mm vs. 4.4 ± 3.0 mm, P <0.05). Also, while birdbeaking was seen in 8 of 20 cases (40%) for the TAG Device, it was only seen in 1 of 12 cases (8%) for the CTAG Device.
CONCLUSION: The present study found enhanced deployment accuracy and conformability along the aortic arch using the CTAG Device compared to the previous-generation TAG Device.

Entities:  

Keywords:  Conformable GORE® TAG® Thoracic Endoprosthesis (CTAG); aortic arch aneurysm; birdbeak; deloyment accuracy; thoracic endovascular aortic repair (TEVAR)

Year:  2015        PMID: 26131025      PMCID: PMC4485048          DOI: 10.3400/avd.oa.14-00141

Source DB:  PubMed          Journal:  Ann Vasc Dis        ISSN: 1881-641X


  6 in total

1.  Influence of endograft oversizing on device migration, endoleak, aneurysm shrinkage, and aortic neck dilation: results from the Zenith Multicenter Trial.

Authors:  W Charles Sternbergh; Samuel R Money; Roy K Greenberg; Timothy A M Chuter
Journal:  J Vasc Surg       Date:  2004-01       Impact factor: 4.268

2.  New C-TAG device and overcome of compression events.

Authors:  K Kasirajan
Journal:  J Cardiovasc Surg (Torino)       Date:  2012-04       Impact factor: 1.888

3.  The impact of bird-beak configuration on aortic remodeling of distal arch pathology after thoracic endovascular aortic repair with the Zenith Pro-Form TX2 thoracic endograft.

Authors:  Hung-Lung Hsu; Chun-Ku Chen; Po-Lin Chen; I-Min Chen; Chiao-Po Hsu; Chih-Wen Chen; Chun-Che Shih
Journal:  J Vasc Surg       Date:  2013-10-16       Impact factor: 4.268

4.  Long-term comparison of thoracic endovascular aortic repair (TEVAR) to open surgery for the treatment of thoracic aortic aneurysms.

Authors:  Nimesh D Desai; Kristen Burtch; William Moser; Pat Moeller; Wilson Y Szeto; Alberto Pochettino; Edward Y Woo; Ronald M Fairman; Joseph E Bavaria
Journal:  J Thorac Cardiovasc Surg       Date:  2012-09       Impact factor: 5.209

Review 5.  Thoracic endovascular aortic repair--indications and evidence.

Authors:  Christopher Q Cao; Paul G Bannon; Rachel Shee; Tristan D Yan
Journal:  Ann Thorac Cardiovasc Surg       Date:  2011       Impact factor: 1.520

6.  Incomplete endograft apposition to the aortic arch: bird-beak configuration increases risk of endoleak formation after thoracic endovascular aortic repair.

Authors:  Takuya Ueda; Dominik Fleischmann; Michael D Dake; Geoffrey D Rubin; Daniel Y Sze
Journal:  Radiology       Date:  2010-05       Impact factor: 11.105

  6 in total
  2 in total

Review 1.  Thoracic Trauma: Aortic Injuries.

Authors:  Akhil Monga; Santosh B Patil; Mathew Cherian; Santhosh Poyyamoli; Pankaj Mehta
Journal:  Semin Intervent Radiol       Date:  2021-04-15       Impact factor: 1.513

2.  Utility of dual longitudinal diameter-reducing ties in aortic arch thoracic endovascular aortic repair.

Authors:  Arkadiusz Kazimierczak; Paweł Rynio; Tomasz Jędrzejczak; Maciej Lewandowski; Piotr Gutowski
Journal:  Postepy Kardiol Interwencyjnej       Date:  2019-12-08       Impact factor: 1.426

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.