Effect of clinical use of nickel-titanium springs.

Amanda Fahning Magno1, André da Costa Monini1, Marisa Veiga Capela2, Lídia Parsekian Martins3, Renato Parsekian Martins4.   

Abstract

INTRODUCTION: Our objectives were to determine whether there are changes on the load deflection rate (L/DP) and the average force (FP) of the superelastic pseudoplateau, and whether permanent deformation is changed in nickel-titanium closed-coil springs (CCSs) after 6 months of clinical use.
METHODS: Twenty-two nickel-titanium CCSs (Sentalloy 100 g; Dentsply GAC, York, Pa) were subjected to tensile mechanical testing at 37°C on activations varying from 3.2 to 16.0 mm before and after 6 months of clinical use. A regression line was fitted over the most horizontal area of the unloading part of the stress-strain graph of every CCS, and its slope was used as L/DP. The FP was determined by the midpoint of the longest segment of the curve that could be fit within the regression line with a R(2) of at least 0.999, and permanent deformation was determined graphically by obtaining the strain value when the measured stress reached zero. The data were analyzed by 3 analyses of variance at 2 levels, with 5% of significance.
RESULTS: Time and activation significantly influenced the variables tested (P < 0.001). Time increased the L/DP and permanent deformation but decreased the FP. Activation decreased L/DP, FP, and permanent deformation. Significant interactions between time and activation were detected for FP (P = 0.013) and deformation (P < 0.001).
CONCLUSIONS: After 6 months of active clinical use, the analyzed springs had a significant but small increase in their L/DP; FP dropped up to 88%, and the CCSs deformed up to 1.26 mm.
Copyright © 2015 American Association of Orthodontists. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2015        PMID: 26124030     DOI: 10.1016/j.ajodo.2015.02.028

Source DB:  PubMed          Journal:  Am J Orthod Dentofacial Orthop        ISSN: 0889-5406            Impact factor:   2.650


  2 in total

1.  Deformation of nickel-titanium closed coil springs: an in vitro study.

Authors:  Camilla Ivini Viana Vieira; José Maurício Dos Santos Nunes Reis; Luiz Geraldo Vaz; Lídia Parsekian Martins; Renato Parsekian Martins
Journal:  Dental Press J Orthod       Date:  2017-02

2.  Effect of immersion in hydrochloric acid and sodium hypochlorite and autoclave sterilization on the force characteristics of orthodontic nickel-titanium open coils.

Authors:  Saeed Noorollahian; Mehrnaz Zakizade
Journal:  Dent Res J (Isfahan)       Date:  2021-02-23
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.